Home / Recalls / Medical Devices
Boston Scientific Neuromodulation Corporation recalls WAVEWRITER ALPHA: Model: SC-1216, 16 IPG KIT, REF: M365SC12160; Model: SC-1232, 32 IPG KIT, REF: M365SC12320
Recalled August 21, 2024 · FDA recall Z-2617-2024 · Boston Scientific Neuromodulation
Hazard
Spinal Cord Stimulation (SCS) Implantable Pulse Generator (IPG) may experience routine system check during IPG charging, which may cause device reset, which may lead to transient loss of stimulation; patients may experience undesired sensations when therapy turns off for approx. 10-15 seconds and then back on, which may lead patient to request surgical intervention for replacement or revision. · FDA Class II
Units
77,674
Sold at
Worldwide Distribution: US (nationwide) including states of: TN, PA, WA, SC, MN, OH, NC, AK, NY, IL, CA, NJ, KY, IA, MO, UT, FL, IN, LA, AL, MS, MD, WI, GA, VA, TX, AZ, KS, OR, CO, CT, ID, AR, OK, MA, MI, ND, NE, SD, WV, NV, HI, NH, DC, MT, ME, VT, WY, DE, NM. OUS (foreign) countries of: Canada, Algeria, Austria, Azerbaijan, Belgium, Bulgaria, Croatia, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Greece, Hungary, Ireland, Israel, Italy, Lebanon, Luxembourg, Netherlands, Norway, Poland, Qatar, Saudi Arabia, Slovakia, Spain, Sweden, Switzerland, Turkey, Ukraine, United Arab Emirates, Portugal, Japan, Australia, New Zealand, Singapore, Argentina, Brazil, Colombia, Puerto Rico
Description
Spinal Cord Stimulation (SCS) Implantable Pulse Generator (IPG) may experience routine system check during IPG charging, which may cause device reset, which may lead to transient loss of stimulation; patients may experience undesired sensations when therapy turns off for approx. 10-15 seconds and then back on, which may lead patient to request surgical intervention for replacement or revision.
Product
WAVEWRITER ALPHA: Model: SC-1216, 16 IPG KIT, REF: M365SC12160; Model: SC-1232, 32 IPG KIT, REF: M365SC12320
Manufacturer(s)
Boston Scientific Neuromodulation Corporation
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →