August 12, 2026Medical DevicesFDA
Boston Scientific Neuromodulation Corporation recalls INFINION CX LEAD KIT, 50CM, Model: SC-2317-50, Material: M365SC2317500; INFINION CX LEAD KIT, 70CM, Model: SC-2317-70,
Hazard: Spinal cord stimulation system lead body may be exposed to stress at the anchor site. This may result in high impedance measurements or lead fractures, which may lead to inadequate stimulation, additional intervention, including lead explant and replacement.
July 15, 2026Medical DevicesFDA
Boston Scientific Neuromodulation Corporation recalls Intracept Access Instruments, Description/REF: INTRACEPT ACCESS INSTRUMENTS- DRILL/RLV0035, INTRACEPT ACCESS INSTRUMEN
Hazard: Intraosseous nerve ablation system instruments may have sterile barrier breach of the outer tray seal, which may introduce a risk of sterility breach, and if breached could lead to severe infection.
June 17, 2026Medical DevicesFDA
Boston Scientific Neuromodulation Corporation recalls Boston Scientific Model SC-4318, Clik X Anchor REF: M365SC43180
Hazard: Sterile anchors my lack sterility assurance.
June 17, 2026Medical DevicesFDA
Boston Scientific Neuromodulation Corporation recalls Boston Scientific TCN Electrodes: REF: TCN-5, TCN Electrode, NITINOL, 5 cm x 0.45 mm (0.0175 in); REF:TCN-10, TCN Elec
Hazard: Reusable electrodes may not meet expected performance levels.
June 17, 2026Medical DevicesFDA
Boston Scientific Neuromodulation Corporation recalls Boston Scientific CSK Electrodes: REF: CSK-TC10, CSK Electrode, STAINLESS STEEL, 10 cm x 28 ga (0.36 mm) REF: CSK-TC10
Hazard: Reusable electrodes may not meet expected performance levels.
September 24, 2025Medical DevicesFDA
Boston Scientific Neuromodulation Corporation recalls Vercise Deep Brain Stimulation Systems Surgical Implant Manual, Document Number: 92328632-09
Hazard: Wire break(s) have occurred in rechargeable deep brain stimulation implantable pulse generators, implanted in the submuscular pectoral location. Wire breaks prevent successful stimulation, thus requiring device removal/replacement and users may experience high monopolar impedances, undesired sensation, loss of therapy, return of symptoms and/or Bluetooth connectivity challenges.
September 10, 2025Medical DevicesFDA
Boston Scientific Neuromodulation Corporation recalls Superion IDS Kit Superion Indirect Decompression System 8MM, VF Implant 101-9808 00884662000529 Superion Indirect
Hazard: As a result of a comprehensive product performance review it was determine that the device and instruments are not consistently meeting expected performance levels. Issues include breakage of the instruments and implant/device.
August 27, 2025Medical DevicesFDA
Boston Scientific Neuromodulation Corporation recalls Burr Hole Cover Kit, REF: M365DB4600C0; Burr Hole Cover Spares Kit, REF: M365DB4605C0. Components of Vercise Deep Stimu
Hazard: Resistance and difficulties could occur while closing the retaining clip locking mechanism, which is part of a burr hole cover kit, which is part of a Deep Brain Stimulation (DBS) system, which may lead to procedural delay, or lead-migration-which may require additional surgery.
August 21, 2024Medical DevicesFDA
Boston Scientific Neuromodulation Corporation recalls WAVEWRITER ALPHA: Model: SC-1216, 16 IPG KIT, REF: M365SC12160; Model: SC-1232, 32 IPG KIT, REF: M365SC12320
Hazard: Spinal Cord Stimulation (SCS) Implantable Pulse Generator (IPG) may experience routine system check during IPG charging, which may cause device reset, which may lead to transient loss of stimulation; patients may experience undesired sensations when therapy turns off for approx. 10-15 seconds and then back on, which may lead patient to request surgical intervention for replacement or revision.
June 5, 2024Medical DevicesFDA
Boston Scientific Neuromodulation Corporation recalls Vercise Genus Deep Brain Stimulation (DBS) Implantable Pulse Generator IPG: R16 IPG KIT, REF: M365DB12160, and R32 IPG K
Hazard: Deep Brain Stimulation (DBS) Implantable Pulse Generator (IPG) may experience routine system check during IPG charging, which may cause device reset. The device reset could lead transient loss of stimulation; patients may experience undesired sensations, transient worsening of movement disorder symptoms, which may lead patient to request surgical intervention for replacement or revision.
June 21, 2023Medical DevicesFDA
Boston Scientific Neuromodulation Corporation recalls Driver Instrument, part of the Superion Indirect Decompression System, REF: 102-9800, used with the Superion Indirect De
Hazard: Pending update to indirect decompression system instructions for use informing users that excessive force during the implant procedure may cause driver instrument tip breaks, which may result in metal fragments (Driver teeth/tips) within the implant location; and if metal fragments are not removed and remain in situ, MRI scans are NOT advised due to potential risk of patient injury.
November 23, 2022Medical DevicesFDA
Boston Scientific Neuromodulation Corporation recalls Superion¿ Indirect Decompression System (IDS) Model Numbers and Size: 101-9808 - 8mm 01-9812 - 12mm 101-9814 - 14m
Hazard: Due to Indirect Decompression System devices exhibiting a potential for weld breakage at the spindle cap-to-main body interface. As a result of the separation of the spindle cap from the main body it may prevent proper deployment of the device during implantation.
June 3, 2020Medical DevicesFDA
Boston Scientific Neuromodulation Corporation recalls Fifty-five (55) cm 8 Contact Extension Kit, part number M365NM3138550 - Product Usage: The Deep Brain Stimulation (DBS)
Hazard: The firm is initiating a voluntary removal of certain spinal cord stimulation leads, lead extensions, adapters, and deep brain stimulation lead extensions due to the "use by date" displayed on the label which is later than the correct "use by date". No adverse health consequence is expected to occur from this issue.
June 3, 2020Medical DevicesFDA
Boston Scientific Neuromodulation Corporation recalls Fifty-five (55) cm 8 Contact Extension Kit, part number M365SC3138550 - Product Usage: Lead Extensions are designed to
Hazard: The firm is initiating a voluntary removal of certain spinal cord stimulation leads, lead extensions, adapters, and deep brain stimulation lead extensions due to the "use by date" displayed on the label which is later than the correct "use by date". No adverse health consequence is expected to occur from this issue.
June 3, 2020Medical DevicesFDA
Boston Scientific Neuromodulation Corporation recalls Precision TM M8 Adapter, 15 cm, part number M365SC9218150 - Product Usage: The adapter is part of a spinal cord stimulat
Hazard: The firm is initiating a voluntary removal of certain spinal cord stimulation leads, lead extensions, adapters, and deep brain stimulation lead extensions due to "use by date" displayed on the label which is later than the correct "use by date". No adverse health consequence is expected to occur from this issue.
June 3, 2020Medical DevicesFDA
Boston Scientific Neuromodulation Corporation recalls The LinearTM 3-4 70 cm 8 Contact Lead Kit, part number M365SC2352700 is part of a spinal cord stimulation system for pr
Hazard: The firm is initiating a voluntary removal of certain spinal cord stimulation leads, lead extensions, adapters, and deep brain stimulation lead extensions due to the "use by date" displayed on the label which is later than the correct "use by date". No adverse health consequence is expected to occur from this issue.
June 3, 2020Medical DevicesFDA
Boston Scientific Neuromodulation Corporation recalls LinearTM 3-4 50 cm 8 Contact Lead Kit, part number M365SC2352500. Leads function as a component of Boston Scientific's
Hazard: The firm is initiating a voluntary removal of certain spinal cord stimulation leads, lead extensions, adapters, and deep brain stimulation lead extensions due to the "use by date" displayed on the label which is later than the correct "use by date". No adverse health consequence is expected to occur from this issue.
June 3, 2020Medical DevicesFDA
Boston Scientific Neuromodulation Corporation recalls Thirty-five (35) cm 8 Contact Extension Kit, part number M365SC3138350. Package Contents: Lead Extension Kit (1) Lead
Hazard: The firm is initiating a voluntary removal of certain spinal cord stimulation leads, lead extensions, adapters, and deep brain stimulation lead extensions due to the "use by date" displayed on the label which is later than the correct "use by date". No adverse health consequence is expected to occur from this issue.
June 3, 2020Medical DevicesFDA
Boston Scientific Neuromodulation Corporation recalls ArtisanTM 70 cm 2 x 8 Surgical Lead Kit, part number M365SC8216700. 2 x 8 Surgical Paddle Lead Kit: (1) One Paddle L
Hazard: The firm is initiating a voluntary removal of certain spinal cord stimulation leads, lead extensions, adapters, and deep brain stimulation lead extensions due to the "use by date" displayed on the label which is later than the correct "use by date". No adverse health consequence is expected to occur from this issue.
April 22, 2020Medical DevicesFDA
Boston Scientific Neuromodulation Corporation recalls Nitinol Thermocouple Electrode, 5cm - Product Usage: The TCN Electrode is indicated for use in RF heat lesioning of peri
Hazard: The firm initiated a voluntary correction of certain Cosman TCN Electrodes due to an incorrect revision of the Instructions for Use (IFU) being packaged with these products.
April 22, 2020Medical DevicesFDA
Boston Scientific Neuromodulation Corporation recalls Nitinol Thermocouple Electrode, 15 cm, 3m Cable - Product Usage: The TCN Electrode is indicated for use in RF heat lesio
Hazard: The firm initiated a voluntary correction of certain Cosman TCN Electrodes due to an incorrect revision of the Instructions for Use (IFU) being packaged with these products.
April 22, 2020Medical DevicesFDA
Boston Scientific Neuromodulation Corporation recalls Nitinol Thermocouple Electrode, 15 cm - Product Usage: The TCN Electrode is indicated for use in RF heat lesioning of pe
Hazard: The firm initiated a voluntary correction of certain Cosman TCN Electrodes due to an incorrect revision of the Instructions for Use (IFU) being packaged with these products.
April 22, 2020Medical DevicesFDA
Boston Scientific Neuromodulation Corporation recalls Nitinol Thermocouple Electrode, 10 cm, 3m Cable - Product Usage: The TCN Electrode is indicated for use in RF heat lesio
Hazard: The firm initiated a voluntary correction of certain Cosman TCN Electrodes due to an incorrect revision of the Instructions for Use (IFU) being packaged with these products.
April 22, 2020Medical DevicesFDA
Boston Scientific Neuromodulation Corporation recalls Nitinol Thermocouple Electrode, 10 cm - Product Usage: The TCN Electrode is indicated for use in RF heat lesioning of pe
Hazard: The firm initiated a voluntary correction of certain Cosman TCN Electrodes due to an incorrect revision of the Instructions for Use (IFU) being packaged with these products.