November 14, 2012Medical DevicesFDA
Zimmer, Inc. recalls MS-30¿ Stem
Hazard: Zimmer Inc., is initiating a correction to the patient labels of products manufactured before March 20I0 by the Zimmer GmbH production site in Winterthur, Switzerland, after receiving four complaints of a discrepancy between product patient labels and product labels with respect to the reference and lot numbers of the product. While the information printed on the product label is correct, the info
November 14, 2012Medical DevicesFDA
Zimmer, Inc. recalls Metasul¿ Durom¿ Femoral component Sterile, Rx
Hazard: Zimmer Inc., is initiating a correction to the patient labels of products manufactured before March 20I0 by the Zimmer GmbH production site in Wintetthur, Switzerland, after receiving four complaints of a discrepancy between product patient labels and product labels with respect to the reference and lot numbers of the product. While the information printed on the product label is correct, the info
November 14, 2012Medical DevicesFDA
Zimmer, Inc. recalls Metasul¿ LDH¿ Head Rx Sterile
Hazard: Zimmer Inc., is initiating a correction to the patient labels of products manufactured before March 20I0 by the Zimmer GmbH production site in Winterthur, Switzerland, after receiving four complaints of a discrepancy between product patient labels and product labels with respect to the reference and lot numbers of the product. While the information printed on the product label is correct, the info
November 14, 2012Medical DevicesFDA
Zimmer, Inc. recalls Alloclassic¿ SL-Offset Stem sterile, Rx
Hazard: Zimmer Inc., is initiating a correction to the patient labels of products manufactured before March 20I0 by the Zimmer GmbH production site in Wintetthur, Switzerland, after receiving four complaints of a discrepancy between product patient labels and product labels with respect to the reference and lot numbers of the product. While the information printed on the product label is correct, the info
November 14, 2012Medical DevicesFDA
Zimmer, Inc. recalls Wagner SL Revision Stem, uncemented Sterile
Hazard: Zimmer Inc., is initiating a correction to the patient labels of products manufactured before March 20I0 by the Zimmer GmbH production site in Winterthur, Switzerland, after receiving four complaints of a discrepancy between product patient labels and product labels with respect to the reference and lot numbers of the product. While the information printed on the product label is correct, the info
November 14, 2012Medical DevicesFDA
Zimmer, Inc. recalls Biolox¿ Option Ceramic Femoral Head System Rx, sterile
Hazard: Zimmer Inc., is initiating a correction to the patient labels of products manufactured before March 20I0 by the Zimmer GmbH production site in Winterthur, Switzerland, after receiving four complaints of a discrepancy between product patient labels and product labels with respect to the reference and lot numbers of the product. While the information printed on the product label is correct, the info
November 14, 2012Medical DevicesFDA
Zimmer, Inc. recalls Biolox¿ delta Ceramic Femoral Head Rx only, sterile
Hazard: Zimmer Inc., is initiating a correction to the patient labels of products manufactured before March 20I0 by the Zimmer GmbH production site in Winterthur, Switzerland, after receiving four complaints of a discrepancy between product patient labels and product labels with respect to the reference and lot numbers of the product. While the information printed on the product label is correct, the info
November 14, 2012Medical DevicesFDA
Zimmer, Inc. recalls Alloclassic SL Shaft
Hazard: Zimmer Inc., is initiating a correction to the patient labels of products manufactured before March 2010 by the Zimmer GmbH production site in Winterthur, Switzerland, after receiving four complaints of a discrepancy between product patient labels and product labels with respect to the reference and lot numbers of the product. While the information printed on the product label is correct, the info
October 24, 2012Medical DevicesFDA
Zimmer, Inc. recalls CPT Femoral Stem 12/14 Neck Taper Extended Offset,(item 00-8114-001-10. CPT stems are primary cemented femoral hip st
Hazard: Zimmer Inc. is initiating a removal of the CPT Modular Stem (item 00-8011-001-00), CPT Stem Petite (item 00-8011-000-05; ) and CPT Femoral Stem 12/14 Neck Taper Extended Offset,(item 00-8114-001-10 ) because devices were packaged in incorrect cartons and distributed internationally and domestically. Zimmer received five complaints of this packaging issue.
October 24, 2012Medical DevicesFDA
Zimmer, Inc. recalls CPT Hip System Femoral Stem Petite, Item No. 00-8011-000-05. CPT stems are primary cemented femoral hip stems used by
Hazard: Zimmer Inc. is initiating a removal of the CPT Modular Stem (item 00-8011-001-00), CPT Stem Petite (item 00-8011-000-05; ) and CPT Femoral Stem 12/14 Neck Taper Extended Offset,(item 00-8114-001-10 ) because devices were packaged in incorrect cartons and distributed internationally and domestically. Zimmer received five complaints of this packaging issue.
October 24, 2012Medical DevicesFDA
Zimmer, Inc. recalls CPT Modular Stem, Item No. 00-8011-001-00; Lot 61947708. CPT stems are primary cemented femoral hip stems used by sur
Hazard: Zimmer Inc. is initiating a removal of the CPT Modular Stem (item 00-8011-001-00), CPT Stem Petite (item 00-8011-000-05; ) and CPT Femoral Stem 12/14 Neck Taper Extended Offset,(item 00-8114-001-10 ) because devices were packaged in incorrect cartons and distributed internationally and domestically. Zimmer received five complaints of this packaging issue.
October 10, 2012Medical DevicesFDA
Zimmer, Inc. recalls Cemented Tibial Drill Product Usage: The cemented tibial drill is used to prepare the tibial bone prior to the impac
Hazard: Zimmer is initiating a recall of the Cemented Tibial Drill for the Zimmer Persona Personalized Knee System due to three reports being received as of August 29. 2012 for tibial cortex perforation during preparation of the tibial bone. The investigation into the reports that have been received found that the perforations are occurring while using the Cemented Tibial Drill. The perforations have o
October 10, 2012Medical DevicesFDA
Zimmer, Inc. recalls 00780304001 MIS 12/14 Slotted Femoral Head Provisional -3.5X36 00780304002 MIS 12/14 Slotted Femoral Head Provisional +
Hazard: Complaints have been received reporting sub-optimal fit between provisional femoral heads (VerSys and MIS) and the mating components (ie stems and provisional femoral necks), in some cases alleging the fit was too loose and in other cases, alleging the fit was too tight. Intraoperative dislodging of the provisional head and the provisional head being left in-vivo after separating from mating comp
October 10, 2012Medical DevicesFDA
Zimmer, Inc. recalls 00780303601 MIS 12/14 Slotted Femoral Head Provisional -3.5X36 00780303602 MIS 12/14 Slotted Femoral Head Provisional +
Hazard: Complaints have been received reporting sub-optimal fit between provisional femoral heads (VerSys and MIS) and the mating components (ie stems and provisional femoral necks), in some cases alleging the fit was too loose and in other cases, alleging the fit was too tight. Intraoperative dislodging of the provisional head and the provisional head being left in-vivo after separating from mating comp
October 10, 2012Medical DevicesFDA
Zimmer, Inc. recalls 00780303201 MIS 12/14 Slotted Femoral Head Provisional -3.5X32 00780303202 MIS 12/14 Slotted Femoral Head Provisional +
Hazard: Complaints have been received reporting sub-optimal fit between provisional femoral heads (VerSys and MIS) and the mating components (ie stems and provisional femoral necks), in some cases alleging the fit was too loose and in other cases, alleging the fit was too tight. Intraoperative dislodging of the provisional head and the provisional head being left in-vivo after separating from mating comp
October 10, 2012Medical DevicesFDA
Zimmer, Inc. recalls 00780302801 MIS 12/14 Slotted Femoral Head Provisional -3.5X28 00780302802 MIS 12/14 Slotted Femoral Head Provisional +
Hazard: Complaints have been received reporting sub-optimal fit between provisional femoral heads (VerSys and MIS) and the mating components (ie stems and provisional femoral necks), in some cases alleging the fit was too loose and in other cases, alleging the fit was too tight. Intraoperative dislodging of the provisional head and the provisional head being left in-vivo after separating from mating comp
October 10, 2012Medical DevicesFDA
Zimmer, Inc. recalls 00780302601 MIS 12/14 Slotted Femoral Head Provisional -3.5X26 00780302602 MIS 12/14 Slotted Femoral Head Provisional +
Hazard: Complaints have been received reporting sub-optimal fit between provisional femoral heads (VerSys and MIS) and the mating components (ie stems and provisional femoral necks), in some cases alleging the fit was too loose and in other cases, alleging the fit was too tight. Intraoperative dislodging of the provisional head and the provisional head being left in-vivo after separating from mating comp
October 10, 2012Medical DevicesFDA
Zimmer, Inc. recalls 00780302201 MIS 12/14 Slotted Femoral Head Provisional -2X22 00780302202 MIS 12/14 Slotted Femoral Head Provisional +0X
Hazard: Complaints have been received reporting sub-optimal fit between provisional femoral heads (VerSys and MIS) and the mating components (ie stems and provisional femoral necks), in some cases alleging the fit was too loose and in other cases, alleging the fit was too tight. Intraoperative dislodging of the provisional head and the provisional head being left in-vivo after separating from mating comp
October 10, 2012Medical DevicesFDA
Zimmer, Inc. recalls 12/14 Neck Taper 00789504001 VerSys Femoral Head Provisional 40mm -3.5 00789504002 VerSys Femoral Head Provisional 40m
Hazard: Complaints have been received reporting sub-optimal fit between provisional femoral heads (VerSys and MIS) and the mating components (ie stems and provisional femoral necks), in some cases alleging the fit was too loose and in other cases, alleging the fit was too tight. Intraoperative dislodging of the provisional head and the provisional head being left in-vivo after separating from mating comp
October 10, 2012Medical DevicesFDA
Zimmer, Inc. recalls 00789503601 VerSys Femoral Head Provisional 36mm -3.5 00789503602 VerSys Femoral Head Provisional 36mm +0 00789503603
Hazard: Complaints have been received reporting sub-optimal fit between provisional femoral heads (VerSys and MIS) and the mating components (ie stems and provisional femoral necks), in some cases alleging the fit was too loose and in other cases, alleging the fit was too tight. Intraoperative dislodging of the provisional head and the provisional head being left in-vivo after separating from mating comp
October 10, 2012Medical DevicesFDA
Zimmer, Inc. recalls 12/14 Neck Taper 00789503201 VerSys Femoral Head Provisional 32mm -3.5 00789503202 VerSys Femoral Head Provisional 32m
Hazard: Complaints have been received reporting sub-optimal fit between provisional femoral heads (VerSys and MIS) and the mating components (ie stems and provisional femoral necks), in some cases alleging the fit was too loose and in other cases, alleging the fit was too tight. Intraoperative dislodging of the provisional head and the provisional head being left in-vivo after separating from mating comp
October 10, 2012Medical DevicesFDA
Zimmer, Inc. recalls 00789502801 VerSys Femoral Head Provisional 28mm -3.5 00789502802 VerSys Femoral Head Provisional 28mm +0 00789502803
Hazard: Complaints have been received reporting sub-optimal fit between provisional femoral heads (VerSys and MIS) and the mating components (ie stems and provisional femoral necks), in some cases alleging the fit was too loose and in other cases, alleging the fit was too tight. Intraoperative dislodging of the provisional head and the provisional head being left in-vivo after separating from mating comp
October 10, 2012Medical DevicesFDA
Zimmer, Inc. recalls 00789502601 VerSys Femoral Head Provisional 26mm -3.5 00789502602 VerSys Femoral Head Provisional 26mm +0 00789502603
Hazard: Complaints have been received reporting sub-optimal fit between provisional femoral heads (VerSys and MIS) and the mating components (ie stems and provisional femoral necks), in some cases alleging the fit was too loose and in other cases, alleging the fit was too tight. Intraoperative dislodging of the provisional head and the provisional head being left in-vivo after separating from mating comp
October 10, 2012Medical DevicesFDA
Zimmer, Inc. recalls 00789502202 VerSys Femoral Head Provisional 22mm +0 00789502203 VerSys Femoral Head Provisional 22mm +3.5 00789502220
Hazard: Complaints have been received reporting sub-optimal fit between provisional femoral heads (VerSys and MIS) and the mating components (ie stems and provisional femoral necks), in some cases alleging the fit was too loose and in other cases, alleging the fit was too tight. Intraoperative dislodging of the provisional head and the provisional head being left in-vivo after separating from mating comp
August 22, 2012Medical DevicesFDA
Zimmer, Inc. recalls 00-4309-029-01 - Trabecular Metal Reverse Shoulder Liner 36 mm 65¿¿ Neck Angle Per surgical technique 97-4309-003-00 Re
Hazard: Lot specific Trabecular Metal (TM) Reverse Shoulder Liner Impactors are being recalled due to reports of alignment peg fractures. All lots of TM Reverse Shoulder Liner Impactors manufactured from 455 stainless steel are being retrieved, as all reported fractured instruments were manufactured with this material. The TM Reverse Shoulder Liner Impactor units that were manufactured with 13-8 stain
August 22, 2012Medical DevicesFDA
Zimmer, Inc. recalls 00-4309-029-00- Trabecular Metal Reverse Shoulder Liner 36mm, Nonsterile Per surgical technique 97-4309-003-00 Revision
Hazard: Lot specific Trabecular Metal (TM) Reverse Shoulder Liner Impactors are being recalled due to reports of alignment peg fractures. All lots of TM Reverse Shoulder Liner Impactors manufactured from 455 stainless steel are being retrieved, as all reported fractured instruments were manufactured with this material. The TM Reverse Shoulder Liner Impactor units that were manufactured with 13-8 stainle
August 22, 2012Medical DevicesFDA
Zimmer, Inc. recalls 00-4309-028-01 - Trabecular Metal Reverse Shoulder Liner 40mm 65¿ Neck Angle Per surgical technique 97-4309-003-00 Revi
Hazard: Lot specific Trabecular Metal (TM) Reverse Shoulder Liner Impactors are being recalled due to reports of alignment peg fractures. All lots of TM Reverse Shoulder Liner Impactors manufactured from 455 stainless steel are being retrieved, as all reported fractured instruments were manufactured with this material. The TM Reverse Shoulder Liner Impactor units that were manufactured with 13-8 stain
August 22, 2012Medical DevicesFDA
Zimmer, Inc. recalls Trabecular Metal Reverse Shoulder Liner 40mm, part 00-4309-028-00 - Per surgical technique 97-4309-003-00 Revision 4
Hazard: Lot specific Trabecular Metal (TM) Reverse Shoulder Liner Impactors are being recalled due to reports of alignment peg fractures. All lots of TM Reverse Shoulder Liner Impactors manufactured from 455 stainless steel are being retrieved, as all reported fractured instruments were manufactured with this material. The TM Reverse Shoulder Liner Impactor units that were manufactured with 13-8 stain
August 22, 2012Medical DevicesFDA
Zimmer, Inc. recalls The NCB Polyaxial Locking Plate System NCB screws 26mm: 02.03155.026, non-sterile Product is implantable screws indic
Hazard: Zimmer investigation following a complaint found one lot of NCB Polyaxial Locking Plate System NCB screws, 26mm, distributed in the United States contained incorrect 32 mm screws.
August 8, 2012Medical DevicesFDA
Zimmer Inc. recalls Nextgen Complete Knee Solution Replacement Jaw for Use with Femoral and Provisional Impactor/Extractor, REF 00-5901-026-
Hazard: The design of the spring clip for the jaw assembly was updated in March 2011 due to fractures of the spring clip. Zimmer has received complaints for the spring clip fracturing on the updated design. The spring clip may fracture or fall off of the instrument during use, and a device fragment may be left in the surgical site. The risk of disassembly or fracture occurs when the instrument is typicall