Home / Recalls / Medical Devices
Zimmer, Inc. recalls 12/14 Neck Taper 00789503201 VerSys Femoral Head Provisional 32mm -3.5 00789503202 VerSys Femoral Head Provisional 32m
Recalled October 10, 2012 · FDA recall Z-2476-2012 · Zimmer
Hazard
Complaints have been received reporting sub-optimal fit between provisional femoral heads (VerSys and MIS) and the mating components (ie stems and provisional femoral necks), in some cases alleging the fit was too loose and in other cases, alleging the fit was too tight. Intraoperative dislodging of the provisional head and the provisional head being left in-vivo after separating from mating comp · FDA Class II
Units
231,641 all lots, all sizes
Sold at
Worldwide Distribution - USA (nationwide) and internationally to Canada, Mexico, South America, Europe, Puerto Rico, China, and Korea.
Description
Complaints have been received reporting sub-optimal fit between provisional femoral heads (VerSys and MIS) and the mating components (ie stems and provisional femoral necks), in some cases alleging the fit was too loose and in other cases, alleging the fit was too tight. Intraoperative dislodging of the provisional head and the provisional head being left in-vivo after separating from mating comp
Product
12/14 Neck Taper 00789503201 VerSys Femoral Head Provisional 32mm -3.5 00789503202 VerSys Femoral Head Provisional 32mm +0 00789503203 VerSys Femoral Head Provisional 32mm +3.5 00789503204 VerSys Femoral Head Provisional 32mm +7 00789503205 VerSys Femoral Head Provisional 32mm +10.5 The provisional heads replicate final femoral implant geometry and are used to trial range of motion, joint stability, and determine appropriate femoral head size and offset during total hip arthroplasty.
Manufacturer(s)
Zimmer, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →