Siemens Healthcare Diagnostics Recalls

692 recalls on record · July 4, 2012 to August 12, 2026 · Medical Devices

Recalls by year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.

September 2, 2015Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc. recalls Dimension Vista B¿2-Microglobulin Flex Reagent Cartridge (B2MIC); Product Usage: The B2MIC method is an in vitro di
Hazard: customers are receiving a high number of Abnormal Assay Errors or are not able to calibrate B2MIC when using Flex reagent cartridge B2MIC various lots. In failure mode, there is the potential for falsely depressed beta2-Microglobulin results on the Dimension Vista system due to assay reagent contamination.
August 26, 2015Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc. recalls ADVIA Chemistry Acetaminophen reagents; for in vitro diagnostic use in the quantitative determination of acetaminophen i
Hazard: A change in the concentration of N-acetylcysteine (NAC) that may cause interference for the ADVIA¿ Chemistry Systems Acetaminophen assay for certain lots.
July 29, 2015Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc recalls ADVIA Centaur Tni-Ultra Assay; SMN 10317708 - 100 tests and SMN 10317709 -500 tests Product Usage: For in vitro diag
Hazard: System-to-system bias between the Tni-Ultra assay on the ADVIA Centaur CP and ADVIA Centaur/ADVIA Centaur XP/ADVIA Centaur XPT systems.
July 15, 2015Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc. recalls Siemens Healthcare Diagnostics Dimension Vista Systems, used with the Dimension Vista Intelligent Lab System software ve
Hazard: Two software issues may occur in Vista software version 3.6.1. Issue #1: Samples may stop processing without notification. Issue #2: Timing issue in Vista 1500 may cause wrong reagent or no reagent delivery. This may lead to unflagged, unexpected low results and results flagged with assay errors.
July 8, 2015Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc recalls The Rapidlab 1265 system is a blood gas analyzer used for laboratory testing of blood gases, electrolytes, metabolites a
Hazard: D50 and D51 diagnostic error codes are not functional.
July 8, 2015Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc recalls The Rapidlab 1260 system is a blood gas analyzer used for laboratory testing of blood gases, electrolytes, metabolites a
Hazard: D50 and D51 Diagnostic error codes are not functional.
July 8, 2015Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc. recalls CentraLink" Data Management System; software is a network solution provider and multi-system data manager for the instru
Hazard: Siemens Healthcare Diagnostics has determined that the sample query function that includes Instrument or Instrument Group as search criteria may not return all samples from the Dimension Vista systems if onboard aliquot support rules are configured in the CentraLink system.
July 8, 2015Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc. recalls EasyLink Data Management System; a clinical data management system that assists medical laboratory professionals with pr
Hazard: System Limitations and Software Issues related to the following features: Custom reference ranges, User-initiated sample and result actions, Auto-verification and quality control rules, and Sample reports.
July 1, 2015Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc. recalls IMMULITE 2000/IMMULITE 2000 XPi Italian Cypress 3g Allergy, Catalog # T23L4 (40 test), Siemens Material Number (SMN) 103
Hazard: IMMULITE 2000/2000XPi Italian Cypress (T23L4), Lots 206 and 208, fail to meet specifications to support the shelf-life stability of 2 years (24 months) expiration date as indicated on the products label.
June 10, 2015Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc. recalls Factor deficient plasma for the Siemens BCS¿ / BCS¿ XP System. In vitro diagnostic reagent for the determination of t
Hazard: Discrepancies for Siemens BCS / BCS XP with dilution studies; erroneously high Factor IX results at the lower end of the measuring range.
June 10, 2015Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc. recalls Dimension Vista Systems LOCI Thyroid Stimulating Hormone; the TSH method is an in vitro diagnostic test for the quantita
Hazard: Dimension Vista LOCI Thyroid Stimulating Hormone (TSH) Lot 14237AA may produce falsely depressed results ranging from -23% to -96% near the end of the well set due to inadequate reagent volume in a Flex well.
June 10, 2015Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc. recalls Dade Actin Activated Cephaloplastin Reagent; Liquid rabbit brain cephalin with plasma activator for use in the determina
Hazard: Recovery for assayed controls are out of assigned ranges. A drift >3 seconds in the normal APTT range and/or >15% in the pathological range has been observed. Patient samples with recoveries close to the medical decision points could exhibit a deviation up to 4 seconds for the normal range and up to 33% for the pathological range.
May 27, 2015Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc. recalls IMMULITE/IMMULITE 1000 Systems TBG, catalog LKTB1 Quantitative measurement of Thyroxine-Binding Globulin (TBG) in ser
Hazard: IMMULITE/IMMULITE 1000 TBG catalog LKTB lot 257 is failing the adjustment slope instrument parameter of <1.80. The resulting slope of the kit is >1.80 resulting in the potential to produce depressed TBG patient values.
May 20, 2015Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc. recalls Dimension EXL with LM, Dimension EXL 200, Dimension RxL, Dimension RxL HM, Dimension RxL Max, Dimension RxL Max HM, Dime
Hazard: Certain Dimension Integrated Chemistry Systems Instrument Power Configurations may have been installed with a non-standard instrument power plug.The use of a non-standard plug configuration as described may lead to a potential electrical hazard. The likelihood of potential clinical impact or injury due to this issue is extremely low.
April 29, 2015Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc. recalls ADVIA Chemistry Lipase Reagent , Catalog Number B01-4840-01 REF: 01984894.
Hazard: Siemens internal investigation confirmed that current contamination avoidance settings do not prevent carryover resulting in falsely elevated ADVIA Chemistry Lipase outlier results when running with Triglycerides, Triglycerides_2, Triglycerides_c, Cholesterol concentrated, and DLDL reagents.
April 15, 2015Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc. recalls ADVIA Chemistry XPT System, Siemens Material Number 10723034, Software Version 1.0.2, Siemens Material Number 11219493,
Hazard: Siemens Healthcare Diagnostics has confirmed an issue with ADVIA Chemistry XPT software version 1.0.2. The system may cause samples to remain in an Inprocess state. Test results on a sample that is held Inprocess will not transmit to the LIS. Manual intervention is necessary to complete the processing of the samples that are held Inprocess.
April 15, 2015Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc. recalls QuikLYTE Dilution Check: used to monitor and adjust for the dilution ratio on Dimension clinical chemistry systems with
Hazard: A positive bias beyond 5% resulting in a dilution check failure or an unnecessary dilution check correction if the positive bias is greater than 1% and less than 5%. The dilution check correction would be detected by QC; however, if not detected by QC, the calibration curve would be decreased and cause elevated sodium, potassium and chloride patient results.
April 1, 2015Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc. recalls ADVIA Chemistry Systems Drug Calibrator I; For in vitro diagnostic use in the calibration of Phenobarbital_2 (PHNB_2), P
Hazard: Incorrect calibrator lot values were assigned for ADVIA Chemistry Systems Drug Calibrator I, lot 3JD018, for the ADVIA Chemistry Systems Phenytoin 2 (PHNY_2) and Phenobarbital 2 (PHNB_2) assays which was the result of calibrator value miscalculation.
March 25, 2015Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc. recalls N Latex IgM; In-vitro diagnostic reagent for the quantitative determination of IgM in human cerebrospinal fluid (CSF) an
Hazard: Complaints have been confirmed on failed proficiency testing due to serum samples recovering too low when using some N Latex IgM reagent kit lots 44018, 44019 and 44042.
March 18, 2015Medical DevicesFDA
Siemens Healthcare Diagnostics Inc recalls Siemens Healthcare Caps for capillary 50/60 uL, Package count 200 For use with blood collection Multicap glass and Mul
Hazard: Capillary Caps for RAPIDLyte Multicap and Multicap-S Blood Collection Capillaries may not attach and result in breakage of the glass capillary
March 18, 2015Medical DevicesFDA
Siemens Healthcare Diagnostics Inc recalls Siemens Healthcare Caps for capillary 140/175 uL, Package count 100 For use with blood collection Multicap glass and
Hazard: Capillary Caps for RAPIDLyte Multicap and Multicap-S Blood Collection Capillaries may not attach and result in breakage of the glass capillary
March 18, 2015Medical DevicesFDA
Siemens Healthcare Diagnostics Inc recalls Siemens Healthcare Caps for capillary 100 uL , Package count 200 For use with blood collection Multicap glass and Multi
Hazard: Capillary Caps for RAPIDLyte Multicap and Multicap-S Blood Collection Capillaries may not attach and result in breakage of the glass capillary
March 18, 2015Medical DevicesFDA
Siemens Healthcare Diagnostics Inc recalls RAPIDLab 1265 System; Siemens Material Number: 10491395, 10321852, 10335524. The Rapidlab 1200 systems are intended
Hazard: Neonatal bilirubin (nBili) parameter may have increased variability when the nBili concentration is greater than 12 mg/dl and the tHb concentration exceeds the upper reportable range of greater than 25 g/dl.
March 18, 2015Medical DevicesFDA
Siemens Healthcare Diagnostics Inc recalls RAPIDLab 1245 System; Siemens Material Number: 10321844, 10491393, 10337179. The Rapidlab 1200 systems are intended
Hazard: Neonatal bilirubin (nBili) parameter may have increased variability when the nBili concentration is greater than 12 mg/dl and the tHb concentration exceeds the upper reportable range of greater than 25 g/dl.
March 18, 2015Medical DevicesFDA
Siemens Healthcare Diagnostics Inc recalls RAPIDPoint 500 Blood Gas Analyzer; Siemens Material Number: 10492730, 10696855, 10696857, 10697306. The RAPIDPoint 5
Hazard: Neonatal bilirubin (nBili) parameter may have increased variability when the nBili concentration is greater than 12 mg/dl and the tHb concentration exceeds the upper reportable range of greater than 25 g/dl.
March 18, 2015Medical DevicesFDA
Siemens Healthcare Diagnostics Inc recalls RAPIDPoint 405 Blood Gas Analyzer; Siemens Material Number: 10310464, 10322347, 10314817, 10317193, 10318999, 10320055,
Hazard: Neonatal bilirubin (nBili) parameter may have increased variability when the nBili concentration is greater than 12 mg/dl and the tHb concentration exceeds the upper reportable range of greater than 25 g/dl.
March 11, 2015Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc recalls ADVIA Centaur Systems Progesterone Kit, (5-pack Ref); In Vitro Diagnostic Catalog number: 02177364; SMN: 10333111.
Hazard: ADVIA Centaur Systems Progesterone Kit Lots Ending in 268 have a high bias on results greater than 30 ng/mL (95.4 nmol/L)
March 11, 2015Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc recalls ADVIA Centaur Systems Progesterone Kit, (5 -pack); In Vitro Diagnostic; Catalog number: 01586287; SMN:10315522.
Hazard: ADVIA Centaur Systems Progesterone Kit Lots Ending in 268 have a high bias on results greater than 30 ng/mL (95.4 nmol/L)
March 11, 2015Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc recalls ADVIA Centaur Systems Progesterone Kit, (1 -pack) In Vitro Diagnostic; Catalog number: 02382928; SMN: 10310305.
Hazard: ADVIA Centaur Systems Progesterone Kit Lots Ending in 268 have a high bias on results greater than 30 ng/mL (95.4 nmol/L)
January 21, 2015Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc. recalls ADVIA Chemistry XPT interfaced to the CentraLink Data Management System V14x in specific configurations: The CentraLink
Hazard: Enabling sending of preliminary/initial results on the ADVIA Chemistry XPT creates a risk that a critical result could be overwritten by the same result rather than showing the repeated result on CentraLink. A result could be erroneous and critical and appear to be verified upon repeat.