Siemens Healthcare Diagnostics, Inc. recalls IMMULITE/IMMULITE 1000 Systems TBG, catalog LKTB1 Quantitative measurement of Thyroxine-Binding Globulin (TBG) in ser

Recalled May 27, 2015 · FDA recall Z-1629-2015 · Siemens Healthcare Diagnostics

Hazard
IMMULITE/IMMULITE 1000 TBG catalog LKTB lot 257 is failing the adjustment slope instrument parameter of <1.80. The resulting slope of the kit is >1.80 resulting in the potential to produce depressed TBG patient values. · FDA Class II
Units
155 kits
Sold at
Worldwide Distribution to the countries of : Denmark, Spain, Great Britain, Germany, Bosnia Herzeg, China and Russian Federation.

Description

IMMULITE/IMMULITE 1000 TBG catalog LKTB lot 257 is failing the adjustment slope instrument parameter of <1.80. The resulting slope of the kit is >1.80 resulting in the potential to produce depressed TBG patient values.

Product
IMMULITE/IMMULITE 1000 Systems TBG, catalog LKTB1 Quantitative measurement of Thyroxine-Binding Globulin (TBG) in serum, as an aid in the diagnosis and treatment of thyroid diseases.
Manufacturer(s)
Siemens Healthcare Diagnostics, Inc.