October 18, 2023Medical DevicesFDA
Encore Medical, LP recalls EMPOWR Acetabular System, Liner, 10¿ Hooded, HXe+, 40H, REF: 942-01-40H
Hazard: Knee insert and acetabular system parts were swapped within inner pack prior to the product being sealed within the tray, which could result in surgeons not having the correct product for the surgery due to different anatomies (Hip vs Knee).
October 18, 2023Medical DevicesFDA
Encore Medical, LP recalls EMPOWR 3D Knee Tibial Insert, 5R, 10MM, VE, REF: 342-10-705
Hazard: Knee insert and acetabular system parts were swapped within inner pack prior to the product being sealed within the tray, which could result in surgeons not having the correct product for the surgery due to different anatomies (Hip vs Knee).
October 18, 2023Medical DevicesFDA
Encore Medical, LP recalls REVERSE PROSTHESIS SHOULDER SYSTEM, SCREW, LOCKING BONE, RSP, 5mmX30mm long, REF: 506-03-130
Hazard: Reverse shoulder prosthesis package that should contain 5mmx30 screws could contain incorrect 3.5mmx30 screws. Use of incorrect screws could result in surgical delay and inadequate fixation.
September 27, 2023Medical DevicesFDA
Encore Medical, LP recalls DJO Surgical: Altivate Reverse, Humeral Stem, Small Shell, Size 6 x 108mm, REF: 533-06-108
Hazard: Orthopedic device components were not subjected to final sterilization before shipment, which if implanted may lead to varying degrees of infection, which may require treatment, prolonged monitoring, and may lead to revision surgery
September 27, 2023Medical DevicesFDA
Encore Medical, LP recalls DJO Surgical: Reverse Shoulder System (RSP) Semi Humeral Socket Insert, 32mm, REF: 508-01-032; Baseplate, Glenoid HA-Coa
Hazard: Orthopedic device components were not subjected to final sterilization before shipment, which if implanted may lead to varying degrees of infection, which may require treatment, prolonged monitoring, and may lead to revision surgery
September 27, 2023Medical DevicesFDA
Encore Medical, LP recalls DJO Surgical: CLP Hip, Stem, Polished Neck, Lateral Offset Implant, Size#3, REF: 426-11-030
Hazard: Orthopedic device components were not subjected to final sterilization before shipment, which if implanted may lead to varying degrees of infection, which may require treatment, prolonged monitoring, and may lead to revision surgery
September 27, 2023Medical DevicesFDA
Encore Medical, LP recalls DJO Surgical: Delta Ceramic Femoral Head, 44 mm + 4.0 mm, REF: 400-03-443
Hazard: Orthopedic device components were not subjected to final sterilization before shipment, which if implanted may lead to varying degrees of infection, which may require treatment, prolonged monitoring, and may lead to revision surgery
September 27, 2023Medical DevicesFDA
Encore Medical, LP recalls DJO Surgical: Empowr Knee, Universal Tibia Baseplate, 6R, REF: 352-02-106
Hazard: Orthopedic device components were not subjected to final sterilization before shipment, which if implanted may lead to varying degrees of infection, which may require treatment, prolonged monitoring, and may lead to revision surgery
April 19, 2023Medical DevicesFDA
Encore Medical, LP recalls EMPOWR Porous Knee Patella Over Drill, REF: 802-05-104; EMPOWR Knee Patella Drill, REF: 802-05-108
Hazard: Patella drills are marked with incorrect diameter size of 5.6mm, when they should be 5.4mm. Porous patella overdrills are marked with incorrect diameter size of 5.4 mm, when they should be 5.6 mm. If the incorrect drill is selected during surgery, there is risk of the implant loosening, implant not properly fitting, and/or fractured bone.
March 8, 2023Medical DevicesFDA
Encore Medical, LP recalls EMPOWR VVC KNEE TIBIAL INSERT, SIZE 6, 22MM, REF: 346-22-706; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 7, 22MM, REF: 346-22-7
Hazard: Tibial insert implant large boxes were labeled small and the small boxes were labeled large. If the incorrect labeling is identified this could lead to surgical delay and if the surgeon does not recognize the size discrepancy between the insert label and the pin label, the surgeon may implant the incorrect pin size.
November 2, 2022Medical DevicesFDA
Encore Medical, LP recalls EMPOWER 3D Knee Tibial Insert, REF: 341-14-708. Orthopedic implant component.
Hazard: Hip liner implant boxes may contain knee inserts and knee insert boxes may contain hip liner implants, which may lead to implants not being available for surgeries, which may cause delays in surgery.
November 2, 2022Medical DevicesFDA
Encore Medical, LP recalls EMPOWR Acetabular System, Liner, 10 deg Hooded, HXe+ 40H, REF: 942-01-40H. Orthopedic implant component.
Hazard: Hip liner implant boxes may contain knee inserts and knee insert boxes may contain hip liner implants, which may lead to implants not being available for surgeries, which may cause delays in surgery.
June 15, 2022Medical DevicesFDA
Encore Medical, LP recalls DJO EMPOWR 3D Knee (Right) 10mm Tibial Insert, REF: 342-10-708
Hazard: A right, size 8, knee tibial insert package may include a mispackaged, left ,size 4, knee tibial insert. If implanted, a risk may include leaving debris in the joint space.
September 1, 2021Medical DevicesFDA
Encore Medical, LP recalls AltiVate Anatomic Shoulder System for the following Part Numbers: 1. Part # 520-08-00, Rev. C., 2. Part # 520-10-000,
Hazard: Complaints have been received concerning the humeral stem failing to mate with the broach handles for a shoulder device. This could result in delay of procedure or revision surgery.
July 28, 2021Medical DevicesFDA
Encore Medical, LP recalls DJO Surgical EMPOWR 3D Knee Tibial Insert - e-plus, REF: 341-16-711 Left Sz. 11 16mm, CE 0086, Sterile H2O2, UDI: (01)00
Hazard: Packaging mixup; There is a potential that packaging identified as 12mm knee prosthesis contains 16mm knee prosthesis and 16mm knee prosthesis packaging contains 12 mm knee prosthesis.
July 28, 2021Medical DevicesFDA
Encore Medical, LP recalls DJO Surgical EMPOWR 3D Knee Tibial Insert - e-plus, REF: 341-12-711, Left Sz. 11 12mm, CE0086, Sterile H2O2, UDI: (01)00
Hazard: Packaging mixup; There is a potential that packaging identified as 12mm knee prosthesis contains 16mm knee prosthesis and 16mm knee prosthesis packaging contains 12 mm knee prosthesis.
May 12, 2021Medical DevicesFDA
Encore Medical, LP recalls DJO EMPOWR KNEE Femoral Sizer
Hazard: Incorrect assembly of a subcomponent by a vendor and as a result, the windows provided to show the size are facing the opposite direction.
December 2, 2020Medical DevicesFDA
Encore Medical, LP recalls EMPOWR Partial Knee Peg Drill; Model: 800-06-008.
Hazard: There have been issues experienced with the peg preparation steps, specifically drilling of the Femoral posterior peg. Upon drilling, the shaft portion of the drill may bind within the Posterior Peg drill guide. The extended length of the drill guide and tight tolerances to ensure accurate preparation seem to be contributing to this situation. There is a potential of a delay in surgery if this issue were to recur.
December 2, 2020Medical DevicesFDA
Encore Medical, LP recalls EMPOWR Partial Knee Peg Drill Guide; Model: 800-06-007.
Hazard: There have been issues experienced with the peg preparation steps, specifically drilling of the Femoral posterior peg. Upon drilling, the shaft portion of the drill may bind within the Posterior Peg drill guide. The extended length of the drill guide and tight tolerances to ensure accurate preparation seem to be contributing to this situation. There is a potential of a delay in surgery if this issue were to recur.
October 28, 2020Medical DevicesFDA
Encore Medical, LP recalls EMPOWR Acetabular System *** EMPOWR Acetabular Cup, Cluster Hole, 54G *** Size 54G 54 mm Dia./Cluster Hole Alpha Code G
Hazard: Due to incomplete machining a small peg remnant on the top of the dome that protrudes above the surrounding porous coating causing the cup not to seat properly.
October 28, 2020Medical DevicesFDA
Encore Medical, LP recalls AltiVate Reverse Small Socket Insert, 44mm Semi Constrained - E-Plus, Part Number: 509-03-044 - Product Usage: Per IFU,
Hazard: Product mislabeled. Use may result in a delay in surgery, stopping of surgery, and revision surgery.
October 14, 2020Medical DevicesFDA
Encore Medical, LP recalls DJO EMPOWR Tibial Knee Punch Handle Part No. 801-05-040 and 801-05-018 contained in Instrument Tray for EMPOWR Porous Kn
Hazard: There is a potential for the retaining pin of the tibial punch handle to become loose and fall out while impacting. If undetected, there is a potential of leaving debris in the joint space.
October 7, 2020Medical DevicesFDA
Encore Medical, LP recalls RSP Standard HUMERAL SOCKET INSERT, 40 MM plus 4MM, HXe-Plus, Part No. 509-00-440
Hazard: Product may be mislabeled with the size etched onto products differing from the size information shown on patient sticker labels. Risks associated with this issue may include the Surgeon implanting the incorrect size; significant physical impairment, discomfort and/or disruption of treatment of underlying condition; and the patient experiencing dislocation resulting in revision surgery.
October 7, 2020Medical DevicesFDA
Encore Medical, LP recalls RSP Standard HUMERAL SOCKET INSERT, 44MM, HXe-Plus, Part No. 509-00-044
Hazard: Product may be mislabeled with the size etched onto products differing from the size information shown on patient sticker labels. Risks associated with this issue may include the Surgeon implanting the incorrect size; significant physical impairment, discomfort and/or disruption of treatment of underlying condition; and the patient experiencing dislocation resulting in revision surgery.
May 20, 2020Medical DevicesFDA
Encore Medical, LP recalls EMPOWR Acetabular System, Non-Sterile, Model Numbers 803-15-023 803-15-026 803-15-027 803-15-028 803-15-029 803-1
Hazard: The firm received complaints that the EMPOWR Acetabular Impactor Balls were fracturing/cracking during impaction which could cause debris to migrate or be left in the joint space.
April 22, 2020Medical DevicesFDA
Encore Medical, LP recalls djosurgical Linear Neck Trials with Spring Sides, Standard, nonsterile, REF S-200635; and djosurgical Linear Neck Trials
Hazard: The retaining ring on the neck trials has a potential to fail interoperatively.
December 18, 2019Medical DevicesFDA
Encore Medical, Lp recalls EMPOWR Locking Femoral Impactor , Part # 800-05-035, packaged in instrument tray FA EMP PAT TOOL KIT Part # 800-99-096
Hazard: The firm has received reports/complaints concerning the breakage of the tips of the locking femoral Impactor. Additionally, if the impactor is not seated correctly on the implant prior to impaction, this may cause off axis loading which can increase the risk of device failure. This could cause a delay in surgery. If the tip is not seen during surgery, long term risks may include an inflammatory response, infection, and/ or pain
August 7, 2019Medical DevicesFDA
Encore Medical, Lp recalls Flexible Drill Shaft, FMP Acetabular System Instrumentation REF 803-05-239 RX Encore Medical, L.P.
Hazard: Manufacturing error, the drill bit shaft is unable to properly capture the drill bit pin. Failure of the drill bit shaft to capture the drill bit pin could result in the drill bit coming loose which could result in a delay in surgery or adverse patient reaction resulting in a revision surgery.
April 17, 2019Medical DevicesFDA
Encore Medical, Lp recalls Exprt Precision System: Revision Hip, Proximal Body with Bolt, Lateral Offset, Sterile R, djo surgical, REF: 495-00-0
Hazard: Complaints regarding loose proximal body bolts or bolts that were found through x-ray as sitting above the proximal body and not properly seated. Patient risks may include: pain, dissociation, and possible revision surgery
September 12, 2018Medical DevicesFDA
Encore Medical, Lp recalls Foundation Modular Femoral Stem, Part Number, 220-10-110. Knee prosthesis component. The Foundation Revision Knee i
Hazard: After receipt of a product complaint for the labeling, it was determined that this device does not have FDA clearance.