Home / Recalls / Medical Devices
Encore Medical, LP recalls DJO EMPOWR 3D Knee (Right) 10mm Tibial Insert, REF: 342-10-708
Recalled June 15, 2022 · FDA recall Z-1257-2022 · Encore Medical
Hazard
A right, size 8, knee tibial insert package may include a mispackaged, left ,size 4, knee tibial insert. If implanted, a risk may include leaving debris in the joint space. · FDA Class II
Units
20
Sold at
U.S. Nationwide distribution in the states of TN, MN, IL, IN, SC, MS, CA, KS.
Description
A right, size 8, knee tibial insert package may include a mispackaged, left ,size 4, knee tibial insert. If implanted, a risk may include leaving debris in the joint space.
Product
DJO EMPOWR 3D Knee (Right) 10mm Tibial Insert, REF: 342-10-708
Manufacturer(s)
Encore Medical, LP
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →