RecallWatchUSA

Home / Recalls / Medical Devices

Zimmer Surgical Inc recalls Zimmer Sterile Disposable Tourniquet Cuff with Protective Sleeve and PLC, Dual Port, Dual Bladder, Item No. 607080152

Recalled March 4, 2020 · FDA recall Z-1356-2020 · Zimmer Surgical

Hazard
When inflating/deflating a single bladder, both bladders could potentially inflate/deflate due to a leak between the bladders. · FDA Class II
Units
58
Sold at
Worldwide distribution - US Nationwide distribution in the states of CA, FL, IA, ID, IL, LA, MA, MO, NC, NY, OH, OR, PA, SC, TN, TX, WA, WI, and WV and country of Netherlands.

Description

When inflating/deflating a single bladder, both bladders could potentially inflate/deflate due to a leak between the bladders.

Product
Zimmer Sterile Disposable Tourniquet Cuff with Protective Sleeve and PLC, Dual Port, Dual Bladder, Item No. 60708015200, UDI (01)00889024377547 (17)230812(10)27968701 The product is a tourniquet - Product Usage: Tourniquets are intended to be used by qualified medical professionals to exert enough pressure on the arterial blood flow in a limb to produce a bloodless operating field.
Manufacturer(s)
Zimmer Surgical Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Zimmer Surgical Inc recalls Zimmer Sterile Disposable Tourniquet Cuff with Protective Sleeve and PLC, Dual Port, Dual Bladder, Item No. 607080152 | RecallWatchUSA