Home / Recalls / Medical Devices
Zimmer, Inc. recalls Trabecular Metal" Reverse Shoulder System
Recalled March 5, 2014 · FDA recall Z-1079-2014 · Zimmer
Hazard
As a result of complaints for disassociation of the TM Reverse Glenosphere from the Baseplate, Zimmer updated the related surgical technique and released a video to provide additional instructions to users on how to properly assemble the Glenosphere to the Baseplate. · FDA Class II
Units
73,002 units
Sold at
Worldwide Distribution-USA (nationwide) and the countries of Germany, India, Malaysia, China, Singapore, Korea, Taiwan, and Brazil.
Description
As a result of complaints for disassociation of the TM Reverse Glenosphere from the Baseplate, Zimmer updated the related surgical technique and released a video to provide additional instructions to users on how to properly assemble the Glenosphere to the Baseplate.
Product
Trabecular Metal" Reverse Shoulder System
Manufacturer(s)
Zimmer, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →