RecallWatchUSA

Home / Recalls / Medical Devices

Zimmer, Inc. recalls PRI Femoral Impactor Block for knee prosthesis surgery. Orthopedic Manual Surgical Instrument

Recalled June 10, 2015 · FDA recall Z-1696-2015 · Zimmer

Hazard
Potential for uncured adhesive between the metal threaded insert and the impactor block to leak out of the assembled part following the autoclave sterilization process. The Master Bond epoxy serves as a secondary locking mechanism to fix the metal insert to the impactor head. Foreign material has been observed on the impactor block. This has the potential for a biological response. · FDA Class II
Units
1,689
Sold at
Worldwide Distribution-US (nationwide) including PR and the states of AL, AZ, CA, CO, FL, IL, IN, KS, MA, MI, MN, MO, NC, NJ, NY, OH, OK, PA, TN, TX, UT, VA, WA, WI, and the countries of Canada, Korea, Singapore, Australia, Japan, Malaysia, India, Austria, Belgium, Canary Islands, Czech Republic, France, Germany, Iceland, Iran, Israel, Italy, Norway, Russian Federation, Saudi Arabia, South America, Spain, Sweden, Switzerland, United Kingdom, and United Arab Emirates.

Description

Potential for uncured adhesive between the metal threaded insert and the impactor block to leak out of the assembled part following the autoclave sterilization process. The Master Bond epoxy serves as a secondary locking mechanism to fix the metal insert to the impactor head. Foreign material has been observed on the impactor block. This has the potential for a biological response.

Product
PRI Femoral Impactor Block for knee prosthesis surgery. Orthopedic Manual Surgical Instrument
Manufacturer(s)
Zimmer, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Zimmer, Inc. recalls PRI Femoral Impactor Block for knee prosthesis surgery. Orthopedic Manual Surgical Instrument | RecallWatchUSA