Home / Recalls / Medical Devices
Wipro GE Healthcare Private Ltd. recalls GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5; Part Number 5867474-04; M
Recalled October 29, 2025 · FDA recall Z-0213-2026 · Wipro GE Healthcare Private
Hazard
Carescape Central Station (CSCS), software version V3.0.5, can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only when the number of central stations connected to the network is 118, 119 or 120. · FDA Class II
Units
173 units
Sold at
Worldwide distribution.
Description
Carescape Central Station (CSCS), software version V3.0.5, can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only when the number of central stations connected to the network is 118, 119 or 120.
Product
GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5; Part Number 5867474-04; Medical Display, cathode ray tube
Manufacturer(s)
Wipro GE Healthcare Private Ltd.
Made in
India
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →