Home / Recalls / Medical Devices
Windstone Medical Packaging, Inc. recalls Monoject 0.9% Sodium Chloride, Flush Prefilled Syringe, 10 mL, REF: 8881570121 inside the following Aligned Medical S
Recalled November 10, 2021 · FDA recall Z-0119-2022 · Windstone Medical Packaging
Hazard
Monoject 0.9% Sodium Chloride Flush Syringe Manufactured by Cardinal Health and placed into kits. Potential for the plunger to draw back after air has been expelled and reintroduce air back into the syringe. · FDA Class I
Units
9,378
Sold at
FL and PA
Description
Monoject 0.9% Sodium Chloride Flush Syringe Manufactured by Cardinal Health and placed into kits. Potential for the plunger to draw back after air has been expelled and reintroduce air back into the syringe.
Product
Monoject 0.9% Sodium Chloride, Flush Prefilled Syringe, 10 mL, REF: 8881570121 inside the following Aligned Medical Solutions Custom Convenience Kits: AMS-9041CP Leaderflex Insertion Kit with Ultrasound, AMS-9046CP-1 Insertion Tray-RX, AMS8939A Universal Procedure Pack w/Split Drape, AMS9957A Port Insertion Pack, and AMS12149 Procedure Pack
Manufacturer(s)
Windstone Medical Packaging, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →