RecallWatchUSA

Home / Recalls / Medical Devices

Wilson-Cook Medical Inc. recalls Cook Medical HEMO-7-EU Hemospray Endoscopic Hemostat, REF G24663, 2.8 cm, sterile.

Recalled December 11, 2024 · FDA recall Z-0582-2025 · Wilson-Cook Medical

Hazard
Specific lots of product have been manufactured with nonconforming raw materials resulting in the red activation knob cracking or breaking at the activation knob internal threading while the device is activated. · FDA Class II
Units
46 devices
Sold at
Distribution was OUS only to Argentina, Brazil, and Israel.

Description

Specific lots of product have been manufactured with nonconforming raw materials resulting in the red activation knob cracking or breaking at the activation knob internal threading while the device is activated.

Product
Cook Medical HEMO-7-EU Hemospray Endoscopic Hemostat, REF G24663, 2.8 cm, sterile.
Manufacturer(s)
Wilson-Cook Medical Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov