Ventana Medical Systems Inc recalls iView DAB Detection Kit, Catalog Number 05266157001, model 760-091 Immunohistochemistry (IHC) for in vitro diagnostic u

Recalled September 26, 2018 · FDA recall Z-2998-2018 · Ventana Medical Systems

Hazard
Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur. · FDA Class I
Units
3068 units
Sold at
worldwide

Description

Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur.

Product
iView DAB Detection Kit, Catalog Number 05266157001, model 760-091 Immunohistochemistry (IHC) for in vitro diagnostic use.
Manufacturer(s)
Ventana Medical Systems Inc