UIH Technologies LLC recalls Shanghai United Imaging Healthcare Co., Ltd. Positron Emission Tomography and Computed Tomography System Model:uMI 55

Recalled October 29, 2025 · FDA recall Z-0223-2026 · UIH Technologies

Hazard
Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement. · FDA Class II
Units
95 units
Sold at
US: OUS:

Description

Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.

Product
Shanghai United Imaging Healthcare Co., Ltd. Positron Emission Tomography and Computed Tomography System Model:uMI 550 MD, Rx Only
Manufacturer(s)
UIH Technologies LLC

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov