Hazard
The arm joints on the television mounting system do not have a washer at the top joint which can result in excessive wear of the stop pins on the arm. This can allow the arm system and items connected to the arm to fall and injure the user or bystanders.
Remedy
Repair
Units
215 distributed to 5 hospitals
Incidents / injuries
The firm received one report of a mounting system falling while being positioned. No injuries have been reported.
Description
The mounting systems were sold for use in hospital Intensive Care Units. The units are off-white and have multiple joints between the ceiling mount and the TV mount at the end of the arm system.
What to do
Secure the unit to prevent movement and contact TRUMPF Medical Systems for information on receiving a repair kit that includes the missing washers.
Contact
For additional information, contact TRUMPF Medical Systems Inc. toll-free at (888) 474-9359 between 8 a.m. and 5 p.m. ET Monday through Friday, or email [email protected]
Product
Ondal AC2000 (PDI) Television Mounting System
Imported by
Ondal Medical Systems of America Inc., of Sandston, VA · TRUMPF Medical Systems Inc., of Charleston, SC
Distributed by
TRUMPF Medical Systems Inc., of Charleston, SC
Retailers
TRUMPF Medical Systems Inc. nationwide From January 2005 through December 2011 for about $2,100.
Made in
Germany