Tosoh Bioscience Inc recalls ST-AIA PACK Fer; Part Number: 025253 Assay, Anemia

Recalled June 13, 2018 · FDA recall Z-2096-2018 · Tosoh Bioscience

Hazard
Asfotase Alfa (Strensiq) interferes with certain Tosoh assays which potentially causes falsely increased or falsely decreased immunoassay test results, depending upon the assay. · FDA Class II
Units
24410
Sold at
Worldwide Distribution: US (Nationwide) and countries of; Canada, Mexico, Columbia, Chile, Uruguay, Nicaragua, Argentina, Peru, Chile, Honduras. Equador, Caymen Islands, Panama, and Costa Rica.

Description

Asfotase Alfa (Strensiq) interferes with certain Tosoh assays which potentially causes falsely increased or falsely decreased immunoassay test results, depending upon the assay.

Product
ST-AIA PACK Fer; Part Number: 025253 Assay, Anemia
Manufacturer(s)
Tosoh Bioscience Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov