Thoratec Corp. recalls Abbott HeartMate Touch Communication System, REF: HMT1100 (US), HMT1100-R (Rental US), HMT1150 (EU)
Recalled December 8, 2021 · FDA recall Z-0308-2022 · Thoratec
Hazard
If LVAS communication system is trying to establish Bluetooth connection and another Bluetooth-enabled device is nearby and advertising for Bluetooth connection, the Bluetooth connectivity interference may cause the system App to close or fail to open. Once Bluetooth adaptor is connected, barring disconnection, all functionality is unaffected; the device will continue to operate as expected. · FDA Class II
Units
988
Sold at
Worldwide distribution - US Nationwide distribution in the states of FL, IL, PA, AR, GA, WI, AZ, TN, TX, MA, NE, CA, VA, NC, OH, MI, DC, IN, NJ, CT, OK, MD, NY, ME, IA, SC, KY, LA, OR, NM, WA, SD, MO, NV, HI, AL, MS, UT, KS and the countries of AUSTRALIA, AUSTRIA, BELGIUM, CANADA, CZECH REPUBLIC, DENMARK, FRANCE, GERMANY, IRELAND, ISRAEL, ITALY, KUWAIT, LITHUANIA, NETHERLANDS, NEW ZEALAND, POLAND, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, UNITED ARAB EMIRATES, UNITED KINGDOM.
Description
If LVAS communication system is trying to establish Bluetooth connection and another Bluetooth-enabled device is nearby and advertising for Bluetooth connection, the Bluetooth connectivity interference may cause the system App to close or fail to open. Once Bluetooth adaptor is connected, barring disconnection, all functionality is unaffected; the device will continue to operate as expected.
Product
Abbott HeartMate Touch Communication System, REF: HMT1100 (US), HMT1100-R (Rental US), HMT1150 (EU)
Manufacturer(s)
Thoratec Corp.