Home / Recalls / Medical Devices
Thor Photomedicine Ltd recalls G041 POD, THOR NOVO GEN 2.0 XL whole body red light therapy bed, model S1177
Recalled December 13, 2023 · FDA recall Z-0454-2024 · Thor Photomedicine
Hazard
Failure of ball stud components causing the gas strut on one side to detach from the canopy, making the canopy difficult to lift. · FDA Class II
Units
6
Sold at
Worldwide Distribution: US (Nationwide): Arizona, California, Colorado, Connecticut, Florida, Hawaii, Idaho, Illinois, Iowa, Maryland, Michigan, Minnesota, Missouri, Nevada, New Hampshire, North Carolina, Ohio, Oregon, Texas, Utah, Virginia, West Virginia. OUS (foreign) to countries of: Australia, Canada, Netherlands, New Zealand, Singapore, United Kingdom.
Description
Failure of ball stud components causing the gas strut on one side to detach from the canopy, making the canopy difficult to lift.
Product
G041 POD, THOR NOVO GEN 2.0 XL whole body red light therapy bed, model S1177
Manufacturer(s)
Thor Photomedicine Ltd
Made in
United Kingdom