RecallWatchUSA

Home / Recalls / Medical Devices

Stryker Neurovascular recalls Stryker brand Neuroform 3 Microdelivery Stent System, 4.5 mm X 20 mm, REF Catalog Number: E345020, Model/Product Number:

Recalled July 31, 2013 · FDA recall Z-1808-2013 · Stryker Neurovascular

Hazard
The device is labeled incorrectly, in that, the device Model Number on the outer carton and inner pouch labels did not match, although both referenced the same lot number. The Model Number on the outer box states UPN M003E345020 (4.5mm x 20 mm Neuroform 3 Stent) while the pouch label states UPN M003E3450300 (4.5mm x 30 mm Neuroform 3 Stent). · FDA Class II
Units
23 units
Sold at
International Distribution including Austria, China, and Germany.

Description

The device is labeled incorrectly, in that, the device Model Number on the outer carton and inner pouch labels did not match, although both referenced the same lot number. The Model Number on the outer box states UPN M003E345020 (4.5mm x 20 mm Neuroform 3 Stent) while the pouch label states UPN M003E3450300 (4.5mm x 30 mm Neuroform 3 Stent).

Product
Stryker brand Neuroform 3 Microdelivery Stent System, 4.5 mm X 20 mm, REF Catalog Number: E345020, Model/Product Number: M003450200, Product is manufactured by Boston Scientific Corporation, Fremont, CA and distributed by Stryker Neurovascular, Fremont, CA The Neuroform Microdelivery Stent System is intended for use with embolic coils for the treatment of wide neck, intracranial, saccular aneurysms arising from a parent vessel with a diameter of greater than equal to 2 mm and greater than equal to 4.5 mm that are not amenable to treatment with surgical clipping.
Manufacturer(s)
Stryker Neurovascular

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Stryker Neurovascular recalls Stryker brand Neuroform 3 Microdelivery Stent System, 4.5 mm X 20 mm, REF Catalog Number: E345020, Model/Product Number: | RecallWatchUSA