Home / Recalls / Medical Devices
Stryker Corporation recalls CinchLock SS Knotless Anchor with Inserter; Model number CAT02462; Orthopedic: The Pivot CinchLock SS Knotless Anc
Recalled May 18, 2016 · FDA recall Z-1634-2016 · Stryker
Hazard
Stryker Endoscopy is recalling the CinchLock SS Knotless Anchor with Inserter and the CinchLock Flex Anchor with Inserter after 13 reports of s of an internal wire breaking upon deployment of the anchor. · FDA Class II
Units
3100 total device, both model numbers.
Sold at
Distributed US (nationwide) and in Australia, Switzerland, Netherlands, Spain and Sweden.
Description
Stryker Endoscopy is recalling the CinchLock SS Knotless Anchor with Inserter and the CinchLock Flex Anchor with Inserter after 13 reports of s of an internal wire breaking upon deployment of the anchor.
Product
CinchLock SS Knotless Anchor with Inserter; Model number CAT02462; Orthopedic: The Pivot CinchLock SS Knotless Anchor (CAT02462) and Pivot CinchLock Flex Knotless Anchor (CAT02463) are non-degradable implant devices intended for the fixation of soft tissue to bone. Used for repair of hip, shoulder, foot and ankle, elbow, wrist and hand and knee.
Manufacturer(s)
Stryker Corporation
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →