RecallWatchUSA

Home / Recalls / Medical Devices

Stryker Corporation recalls CinchLock Flex Knotless Anchor with Inserter; Model number CAT02643; Orthopedic: The Pivot CinchLock SS Knotless An

Recalled May 18, 2016 · FDA recall Z-1635-2016 · Stryker

Hazard
Stryker Endoscopy is recalling the CinchLock SS Knotless Anchor with Inserter and the CinchLock Flex Anchor with Inserter after 13 reports of an internal wire breaking upon deployment of the anchor. · FDA Class II
Units
3100 total devices, both models
Sold at
Distributed US (nationwide) and in Australia, Switzerland, Netherlands, Spain and Sweden.

Description

Stryker Endoscopy is recalling the CinchLock SS Knotless Anchor with Inserter and the CinchLock Flex Anchor with Inserter after 13 reports of an internal wire breaking upon deployment of the anchor.

Product
CinchLock Flex Knotless Anchor with Inserter; Model number CAT02643; Orthopedic: The Pivot CinchLock SS Knotless Anchor (CAT02462) and Pivot CinchLock Flex Knotless Anchor (CAT02463) are non-degradable implant devices intended for the fixation of soft tissue to bone. Used for repair of hip, shoulder, foot and ankle, elbow, wrist and hand and knee.
Manufacturer(s)
Stryker Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Stryker Corporation recalls CinchLock Flex Knotless Anchor with Inserter; Model number CAT02643; Orthopedic: The Pivot CinchLock SS Knotless An | RecallWatchUSA