RecallWatchUSA

Home / Recalls / Medical Devices

Steris Corporation recalls VERIFY Steam Test Pack (20 Tests and 5 controls) Item Number: S3065 Intended for use in installation testing and rou

Recalled January 11, 2023 · FDA recall Z-0780-2023 · Steris

Hazard
Biological indicator (BI) used to promote growth of the G. stearothermophilus and B. atrophaeus spores was inconsistent in promoting growth throughout lot , some BIs within the lot performed as intended while others did not. When used to monitor steam and ethylene oxide sterilization cycles, the potential hazard is that the BI will show no growth indicating that the cycle was effective, even if it may not have been. · FDA Class II
Units
60 boxes
Sold at
Nationwide Foreign: AE AT CA CO ES FI GU IN IT JP MX SG TH TW ZA

Description

Biological indicator (BI) used to promote growth of the G. stearothermophilus and B. atrophaeus spores was inconsistent in promoting growth throughout lot , some BIs within the lot performed as intended while others did not. When used to monitor steam and ethylene oxide sterilization cycles, the potential hazard is that the BI will show no growth indicating that the cycle was effective, even if it may not have been.

Product
VERIFY Steam Test Pack (20 Tests and 5 controls) Item Number: S3065 Intended for use in installation testing and routine monitoring of steam and ethylene oxide sterilization processes
Manufacturer(s)
Steris Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov