Home / Recalls / Medical Devices
Steris Corporation recalls Eagle 3000 Sterilizer, small steam sterilizer. Product Usage: The Eagle 3000 Small Sterilizer is designed for sterilizat
Recalled December 11, 2019 · FDA recall Z-0604-2020 · Steris
Hazard
Potential for the incorrect pressure safety valve being included in the Preventive Maintenance (PM) kit for the 3000 Small Sterilizers. · FDA Class II
Units
236
Sold at
Worldwide distribution - US Nationwide distribution in the states of AL AR AZ CA CO DE FL GA HI IA ID IL IN KS LA MD ME MI MN MO MS MT NC NE NH NJ NM NV NY OH OK OR PA RI SC SD TN TX UT VA WA WI WV, and country of Canada.
Description
Potential for the incorrect pressure safety valve being included in the Preventive Maintenance (PM) kit for the 3000 Small Sterilizers.
Product
Eagle 3000 Sterilizer, small steam sterilizer. Product Usage: The Eagle 3000 Small Sterilizer is designed for sterilization of certain materials used in healthcare facilities, and are available in two configurations: Gravity designed for sterilization of non-porous, heat- and moisture stabile goods, and sterilization of items intended for immediate use. Prevacuum designed for fast, efficient sterilization of porous, heat- and moisture-stabile materials, in addition to the same sterilization capabilities as a Gravity sterilizer.
Manufacturer(s)
Steris Corporation
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →