Steris Corporation recalls AMSCO¿ Evolution¿ steam sterilizer Transfer Carriage. Product Usage: Mobile, wheeled transportation unit utilized t

Recalled December 18, 2013 · FDA recall Z-0478-2014 · Steris

Hazard
The firm discovered that unlatching can occur when overloaded transfer carriages are pulled or maneuvered incorrectly or with excessive force. · FDA Class II
Units
936 units
Sold at
Worldwide Distribution - USA Nationwide in the states of: AL, AZ, AR, CA, CO, CT, DE, FL, GA, ID, IL, IN, KS, KY, LA, ME, MD, MA, MI, MN, MI, MO, MT, NE, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, WY and the District of Columbia including the countries of: Brazil, Canada, Chile, Guam, Mexico, Saudi Arabia, United Arab Emirates & Venezuela.

Description

The firm discovered that unlatching can occur when overloaded transfer carriages are pulled or maneuvered incorrectly or with excessive force.

Product
AMSCO¿ Evolution¿ steam sterilizer Transfer Carriage. Product Usage: Mobile, wheeled transportation unit utilized to carry, load, & unload the Loading Car into and out of steam sterilizer units.
Manufacturer(s)
Steris Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov