RecallWatchUSA

Home / Recalls / Medical Devices

Skytron, Div. The KMW Group, Inc recalls EZ Slide Battery/line powered hydraulic table with removable leg section and 21 top slide General Purpose Surgical Ta

Recalled June 8, 2016 · FDA recall Z-1844-2016 · Skytron, Div. The KMW Group

Hazard
The potential exists within the identified tables that incorrectly manufactured parts have been used in assembly. If a table is found to have these incorrectly manufactured parts, it is possible for the tables elevation column slide cap to become detached from the elevation column and could result in the loss of control of the table top and injury to the patient or staff. · FDA Class II
Units
458
Sold at
Nationwide Distribution to AL, AZ, CO, IA, WI, PA, TN, MA, TX, MI, FL, MS, LA, and CA.

Description

The potential exists within the identified tables that incorrectly manufactured parts have been used in assembly. If a table is found to have these incorrectly manufactured parts, it is possible for the tables elevation column slide cap to become detached from the elevation column and could result in the loss of control of the table top and injury to the patient or staff.

Product
EZ Slide Battery/line powered hydraulic table with removable leg section and 21 top slide General Purpose Surgical Table Skytron EZ Slide Operating Table General Purpose Surgical Table
Manufacturer(s)
Skytron, Div. The KMW Group, Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov