Siemens Medical Solutions USA, Inc recalls ONCOR Impression, ONCOR Impression plus, and ONCOR Expression Medical charged-particle radiation therapy system Produ
Recalled February 22, 2017 · FDA recall Z-1214-2017 · Siemens Medical Solutions USA
Hazard
Siemens became aware of Incorrect values for the rotational collimator position displayed due to mechanical problems of the sensor slide. In addition, there are improper weld seam at the overhead suspension. · FDA Class II
Units
194 systems
Sold at
United States distribution
Description
Siemens became aware of Incorrect values for the rotational collimator position displayed due to mechanical problems of the sensor slide. In addition, there are improper weld seam at the overhead suspension.
Product
ONCOR Impression, ONCOR Impression plus, and ONCOR Expression Medical charged-particle radiation therapy system Product Usage: The intended use of the SIEMENS branded PRIMUS family of linear accelerator systems is to deliver x-ray radiation for therapeutic treatment of cancer.