RecallWatchUSA

Home / Recalls / Medical Devices

Siemens Medical Solutions USA, Inc recalls MEVATRON M2/Primus Mid-Energy Model #01940035, MEVATRON PRIMART #05500371, MEVATRON KD-2 #08515520, MEVATRON KD2 #019407

Recalled April 15, 2020 · FDA recall Z-1678-2020 · Siemens Medical Solutions USA

Hazard
Potential safety issue with the dose monitoring system safety interlocks which could result in a 10-20% overdose of radiation therapy. · FDA Class II
Units
1012
Sold at
Worldwide distribution. US consignees in the following states: Alabama, California, Delaware, Florida, Georgia, Iowa, Illinois, Kentucky, Louisiana, Massachusetts, Maryland, Maine, Michigan, Minnesota, Missouri, Mississippi, North Carolina, Nebraska, New Jersey, Nevada, New York, Ohio, Oklahoma, Pennsylvania, Puerto Rico, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Wisconsin, West Virginia, Wyoming

Description

Potential safety issue with the dose monitoring system safety interlocks which could result in a 10-20% overdose of radiation therapy.

Product
MEVATRON M2/Primus Mid-Energy Model #01940035, MEVATRON PRIMART #05500371, MEVATRON KD-2 #08515520, MEVATRON KD2 #01940753, MEVATRON M-2 6740 #09401506, MEVATRON MD-2 #09401654, MEVATRON MDX-2 #09401746 with Digital LINAC Systems Control Console Software Versions 6, 7, 9, 13
Manufacturer(s)
Siemens Medical Solutions USA, Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Siemens Medical Solutions USA, Inc recalls MEVATRON M2/Primus Mid-Energy Model #01940035, MEVATRON PRIMART #05500371, MEVATRON KD-2 #08515520, MEVATRON KD2 #019407 | RecallWatchUSA