Siemens Medical Solutions USA, Inc recalls Artis zeego (Model no. 10280959) and Artis Q.zeego (Model no. 10848283), Interventional Fluoroscopic X-Ray System

Recalled July 13, 2022 · FDA recall Z-1343-2022 · Siemens Medical Solutions USA

Hazard
It may occur that after system startup no stand movement is possible any longer in the event of a discharged BIOS battery of the robotic stand control PC. If this problem occurs during startup, all stand movements are blocked and can only be reactivated by a field service engineer. · FDA Class II
Units
233 worldwide; 32 US
Sold at
Worldwide distribution - US Nationwide distribution in the states of AL, AR, CA, CO, FL, IL, LA, ME, MN, MO, NC, OH, PA, SC, TN, TX, WI and the countries of Australia, Austria, Azerbaijan, Belgium, Canada, China, Finland, France, Germany, Hong Kong, India, Italy, Japan, Lebanon, Lithuania, Luxembourg, Netherlands, Poland, Portugal, South Korea, Switzerland, United Kingdom.

Description

It may occur that after system startup no stand movement is possible any longer in the event of a discharged BIOS battery of the robotic stand control PC. If this problem occurs during startup, all stand movements are blocked and can only be reactivated by a field service engineer.

Product
Artis zeego (Model no. 10280959) and Artis Q.zeego (Model no. 10848283), Interventional Fluoroscopic X-Ray System
Manufacturer(s)
Siemens Medical Solutions USA, Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov