Siemens Medical Solutions USA, Inc recalls Artis zee/ zeego systems; Product Usage: Usage: Artis zee/ zeego systems are dedicated angiography systems develop
Recalled July 1, 2015 · FDA recall Z-1846-2015 · Siemens Medical Solutions USA
Hazard
There exists a possible position sensor fault in the swivel base axis not being detected by the system software. When moving, the system could potentially exceed its usual speed, resulting in a collision. No injuries reported. · FDA Class II
Units
475
Sold at
US Nationwide Distribution
Description
There exists a possible position sensor fault in the swivel base axis not being detected by the system software. When moving, the system could potentially exceed its usual speed, resulting in a collision. No injuries reported.
Product
Artis zee/ zeego systems; Product Usage: Usage: Artis zee/ zeego systems are dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed include cardiac angiography, neuro angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. .