Siemens Medical Solutions USA, Inc recalls Artis zee biplane, Model Number 10094141
Recalled July 31, 2019 · FDA recall Z-2031-2019 · Siemens Medical Solutions USA
Hazard
A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure. · FDA Class II
Units
268
Sold at
US Nationwide Distribution
Description
A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.
Product
Artis zee biplane, Model Number 10094141
Manufacturer(s)
Siemens Medical Solutions USA, Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →