RecallWatchUSA

Home / Recalls / Medical Devices

Siemens Medical Solutions USA, Inc recalls ARTIS Pheno VE30A and VE40A, Model 10849000

Recalled April 22, 2026 · FDA recall Z-1814-2026 · Siemens Medical Solutions USA

Hazard
During 3D acquisitions, lighter and darker patient images may be captured, which may result in less accurate 3D reconstruction. The variation of brightness is a result of a constant unregulated medium dose and the angular change of patient diameter caused by rotational acquisition. This may result in unintentional low-dose radiation exposure to a patient. · FDA Class II
Units
73
Sold at
U.S. and O.U.S.

Description

During 3D acquisitions, lighter and darker patient images may be captured, which may result in less accurate 3D reconstruction. The variation of brightness is a result of a constant unregulated medium dose and the angular change of patient diameter caused by rotational acquisition. This may result in unintentional low-dose radiation exposure to a patient.

Product
ARTIS Pheno VE30A and VE40A, Model 10849000
Manufacturer(s)
Siemens Medical Solutions USA, Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov