Siemens Medical Solutions USA, Inc recalls Artis one fluoroscopic x-ray system and patient table Artis one is an angiography system developed for diagnostic i

Recalled March 28, 2018 · FDA recall Z-1148-2018 · Siemens Medical Solutions USA

Hazard
Certain patient table systems may have a missing or incorrectly installed snap ring that could cause the bolt to slide out of its base and the tabletop carriage to slide to the floor. · FDA Class II
Units
16
Sold at
US Distribution to the states of : AL, AR, CA, FL, IN, LA, MI, MO, NC, NY, PA, TX, and WA.

Description

Certain patient table systems may have a missing or incorrectly installed snap ring that could cause the bolt to slide out of its base and the tabletop carriage to slide to the floor.

Product
Artis one fluoroscopic x-ray system and patient table Artis one is an angiography system developed for diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Arts one include cardiac angiography, neuro-angiography, general angiography, rotational angiography; multipurpose angiography and whole body radiographic/fluoroscopic procedures as well as procedures next to the table for i.e. patient extremities.
Manufacturer(s)
Siemens Medical Solutions USA, Inc