Siemens Medical Solutions USA, Inc. recalls ACUSON SC2000 Ultrasound System. The firm name on the label is Siemens Medical Solutions USA, Inc., Buffalo Grove, IL.

Recalled May 16, 2018 · FDA recall Z-1200-2018 · Siemens Medical Solutions USA

Hazard
The ECG signal may flatline due to electromagnetic interference during the use of electrosurgical equipment. · FDA Class II
Units
229 systems
Sold at
Distribution was made to medical facilities in AZ, CA, CT, FL, IL, IN, KS, KY, MI, MN, MO, NJ, NY, NC, OH, PA, TX, VA, and WA. There was no military/government distribution. Foreign distribution was made to Argentina, Australia, Austria, Belgium, Bosnia Herzeg, Colombia, Croatia, Denmark, France, Germany, India, Ireland, Israel, Italy, Japan, Mexico, New Zealand, Philippines, Poland, Republic Korea, Romania, Singapore, Slovakia, Spain, Sweden, Taiwan, Thailand, Turkey, UAE, and Vietnam.

Description

The ECG signal may flatline due to electromagnetic interference during the use of electrosurgical equipment.

Product
ACUSON SC2000 Ultrasound System. The firm name on the label is Siemens Medical Solutions USA, Inc., Buffalo Grove, IL.
Manufacturer(s)
Siemens Medical Solutions USA, Inc.