Siemens Healthcare Diagnostics, Inc. recalls The ADVIA Chemistry XPT Chemistry System, Siemens Material Number (SMN) 10723034, is an automated, clinical chemistry an

Recalled July 14, 2021 · FDA recall Z-2014-2021 · Siemens Healthcare Diagnostics

Hazard
ADVIA Chemistry System and Software Versions (V1.4 SMN 11314625 and V1.4 China SMN 11316885) may not be aligned with customer expectations for the current default mixer failure detection setting of 2; an erroneously elevated or depressed patient result to be reported without alert on the test result · FDA Class II
Units
842 units
Sold at
Worldwide distribution - US Nationwide distribution in the states of CA, CO, CT, FL, MA, MD, MI, MN, NY, OH, PA, TX and the countries of Australia, Bahrain, Brazil, Canada, Chile, Croatia, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, India, Italy, Malaysia, Netherlands, New Zealand, Norway, P.R. China, Portugal, Republic Korea, Saudi Arabia, Slovakia, Spain, Sweden, Taiwan, Thailand, Turkey, United Kingdom, Vietnam.

Description

ADVIA Chemistry System and Software Versions (V1.4 SMN 11314625 and V1.4 China SMN 11316885) may not be aligned with customer expectations for the current default mixer failure detection setting of 2; an erroneously elevated or depressed patient result to be reported without alert on the test result

Product
The ADVIA Chemistry XPT Chemistry System, Siemens Material Number (SMN) 10723034, is an automated, clinical chemistry analyzer that runs tests on serum, plasma, urine, or cerebral spinal fluid in random access and batch modes. Tests performed using this system are intended for in vitro diagnostic use.
Manufacturer(s)
Siemens Healthcare Diagnostics, Inc.