RecallWatchUSA

Home / Recalls / Medical Devices

Siemens Healthcare Diagnostics, Inc recalls Siemens Healthcare ADVIA Centaur DHEA-504 (DHEAS) 50 Test 06489701 The ADVIA Centaur DHEAS assay is an in vitro diag

Recalled December 19, 2012 · FDA recall Z-0509-2013 · Siemens Healthcare Diagnostics

Hazard
Operating range of the ADVIA Centaur systems is 18C to 30¿C (64.4¿F to 86.0¿F), a change in room temperature may increase or decrease assay results, depending on the specific assay and ADVIA Centaur system used. · FDA Class II
Units
12976
Sold at
Worldwide Distribution - USA (nationwide) and Internationally to the following countries: Argentina Australia Austria Azerbaijan Belgium Bosnia Herzeg. Brazil Canada Chile COLOMBIA Croatia Czech Republic Denmark Egypt Estonia Finland France Germany Greece Hong Kong Hungary Iceland India Indonesia Iran Ireland Italy Japan Kazakhstan Latvia Lithuania Malaysia Mexico Netherlands New Caledonia Norway Oman P.R. China Philippines Poland Portugal Qatar Rep. of Yemen Republic Korea Reunion Romania Russian Fed. Saudi Arabia Serbia Slovakia Slovenia South Africa Spain Sweden Switzerland Taiwan Thailand Turkey U.A.E. United Kingdom Uruguay Vietnam

Description

Operating range of the ADVIA Centaur systems is 18C to 30¿C (64.4¿F to 86.0¿F), a change in room temperature may increase or decrease assay results, depending on the specific assay and ADVIA Centaur system used.

Product
Siemens Healthcare ADVIA Centaur DHEA-504 (DHEAS) 50 Test 06489701 The ADVIA Centaur DHEAS assay is an in vitro diagnostic immunoassay for the quantitative determination of dehydroepiandrosterone sulfate (DHEAS) in human serum and plasma using the ADVIA Centaur and ADVIA Centaur XP systems
Manufacturer(s)
Siemens Healthcare Diagnostics, Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Siemens Healthcare Diagnostics, Inc recalls Siemens Healthcare ADVIA Centaur DHEA-504 (DHEAS) 50 Test 06489701 The ADVIA Centaur DHEAS assay is an in vitro diag | RecallWatchUSA