Siemens Healthcare Diagnostics, Inc. recalls Siemens Dimension Vista(R) System B12 Flex(R) Reagent Cartridge (K6442) B12 Flex Reagent Cartridge used with the Dime
Recalled July 4, 2012 · FDA recall Z-1869-2012 · Siemens Healthcare Diagnostics
Hazard
Siemens Healthcare Diagnostics has confirmed that the presence of Intrinsic Factor Blocking Antibody in some patients can result in a false and significant elevation in the reported concentration of Vitamin B12 on the Dimension Vista System. This issue affects all lots of B12 Flex reagent cartridges on the Dimension Vista System. · FDA Class II
Units
59,735
Sold at
Worldwide Distribution -- USA (nationwide) and the countries of Austria, Belgium, Czech Republic, Denmark, Finland, France, Germany, Great Britian, Italy, Netherlands, Norway, Portugal, Slovakia, South Africa, Spain, Switzerland, United Arab Emirates, Canada, New Zealand, and South Korea. *** Center Recommended Depth is Retail***
Description
Siemens Healthcare Diagnostics has confirmed that the presence of Intrinsic Factor Blocking Antibody in some patients can result in a false and significant elevation in the reported concentration of Vitamin B12 on the Dimension Vista System. This issue affects all lots of B12 Flex reagent cartridges on the Dimension Vista System.
Product
Siemens Dimension Vista(R) System B12 Flex(R) Reagent Cartridge (K6442) B12 Flex Reagent Cartridge used with the Dimension Vista System for the in vitro diagnostic test for the quantitative measurement of Vitamin B12 in human serum and plasma.
Manufacturer(s)
Siemens Healthcare Diagnostics, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →