Siemens Healthcare Diagnostics, Inc. recalls siemens Dimension¿ Phenytoin Flex¿ Reagent Cartridge The PTN method used on the Dimension¿ clinical chemistry system

Recalled November 5, 2014 · FDA recall Z-0148-2015 · Siemens Healthcare Diagnostics

Hazard
Internal testing has confirmed a low absorbance range between the levels 1 and 5 calibrators with this lot resulting in the potential for a positive or negative bias and imprecision across the PTN assay range. · FDA Class II
Units
1733
Sold at
Worldwide Distribution - USA (nationwide) including Puerto Rico., and the countries of Australia, Brazil, Bahamas, Canada, Chile, India, Japan and Uruguay

Description

Internal testing has confirmed a low absorbance range between the levels 1 and 5 calibrators with this lot resulting in the potential for a positive or negative bias and imprecision across the PTN assay range.

Product
siemens Dimension¿ Phenytoin Flex¿ Reagent Cartridge The PTN method used on the Dimension¿ clinical chemistry system is an in vitro diagnostic test intended to measure phenytoin (dilantin, diphenylhydantoin), an anti-epileptic drug, in human serum and plasma. PTN test results are used in the diagnosis and treatment of phenytoin overdose and in monitoring levels of phenytoin to ensure appropriate therapy.