Siemens Healthcare Diagnostics, Inc. recalls Siemens Dimension LOCI TNI Troponin I Calibrator RC 621. The LOCI TNI Troponin I Calibrator is an in vitro diagno

Recalled July 2, 2014 · FDA recall Z-1926-2014 · Siemens Healthcare Diagnostics

Hazard
Siemens Dimension LOCI TNI Troponin I calibrator lot 3DD007 may demonstrate an average upward shift of 24% in QC and patient results following calibration. · FDA Class II
Units
1625
Sold at
Distribution: US Nationwide including Puerto Rico

Description

Siemens Dimension LOCI TNI Troponin I calibrator lot 3DD007 may demonstrate an average upward shift of 24% in QC and patient results following calibration.

Product
Siemens Dimension LOCI TNI Troponin I Calibrator RC 621. The LOCI TNI Troponin I Calibrator is an in vitro diagnostic product for the calibration of the cardiac troponin I (TNI) method for the Dimension clinical EXL integrated chemistry system with LOCI module.
Manufacturer(s)
Siemens Healthcare Diagnostics, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov