Siemens Healthcare Diagnostics Inc recalls RP500e Handheld Barcode Scanner, Zebra Model: DS4308 - HC0062BZZWW, Siemens Material Number (SMN) 11416778

Recalled July 1, 2020 · FDA recall Z-2432-2020 · Siemens Healthcare Diagnostics

Hazard
The scanner is not confirming the integrity of read barcode data using the check-digit when reading Code 39 barcodes. · FDA Class II
Units
436
Sold at
Foreign distribution only.

Description

The scanner is not confirming the integrity of read barcode data using the check-digit when reading Code 39 barcodes.

Product
RP500e Handheld Barcode Scanner, Zebra Model: DS4308 - HC0062BZZWW, Siemens Material Number (SMN) 11416778
Manufacturer(s)
Siemens Healthcare Diagnostics Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov