RecallWatchUSA

Home / Recalls / Medical Devices

Siemens Healthcare Diagnostics, Inc recalls IMMULITE Systems Third Generation TSH (REF LKTS1 - 100T, LKTS5 - 500T, L2KTS2 - 200T, L2KTS6 - 600T). Intended to qua

Recalled August 14, 2013 · FDA recall Z-1885-2013 · Siemens Healthcare Diagnostics

Hazard
A rare variant of TSH, identified in a small cluster of patients, is not detected. · FDA Class II
Units
51792
Sold at
Worldwide Distribution-USA (nationwide) and the countries of Canada, AE, AR, AT, AU, BA, BD, BE, BG, BO, BR, BY, CH, CL, CM, CN, CO, CR, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FR, GB, GR, GT, HK, HN, HR, HU, ID, IE, IL, IN, IR, IS, IT, JO, JP, KA, KR, KW, KZ, LB, LK, LT, LV, LY, MA, MK, MT, MX, MY, NI, NL, NO, NP, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RU, SA, SE, SG, SK, SV, SY, TH, TM, TN, TR, TW, UA, UY, UZ, VE, VN, XS, and ZA.

Description

A rare variant of TSH, identified in a small cluster of patients, is not detected.

Product
IMMULITE Systems Third Generation TSH (REF LKTS1 - 100T, LKTS5 - 500T, L2KTS2 - 200T, L2KTS6 - 600T). Intended to quantitatively measure human thyroid stimulating hormone (TSH, thyrotropin).
Manufacturer(s)
Siemens Healthcare Diagnostics, Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov