Siemens Healthcare Diagnostics, Inc. recalls Dimension Vista(R) 500 and Dimension Vista(R) 1500 - Reloaded Vista Vials - limited to systems using software versions 3

Recalled December 11, 2013 · FDA recall Z-0430-2014 · Siemens Healthcare Diagnostics

Hazard
There is the potential for the vial volume of "in use" Vista calibrators, controls, or sample diluent to incorrectly revert to full volume when removed and reloaded. Dimension Vista(R) 500 and Dimension Vista(R) 1500 - Reloaded Vista Vials - limited to systems using software versions 3.4 and 3.5.1 · FDA Class II
Units
776
Sold at
Worldwide Distribution: USA (nationwide) and iinternationally to Canada.

Description

There is the potential for the vial volume of "in use" Vista calibrators, controls, or sample diluent to incorrectly revert to full volume when removed and reloaded. Dimension Vista(R) 500 and Dimension Vista(R) 1500 - Reloaded Vista Vials - limited to systems using software versions 3.4 and 3.5.1

Product
Dimension Vista(R) 500 and Dimension Vista(R) 1500 - Reloaded Vista Vials - limited to systems using software versions 3.4 and 3.5.1. Diagnostic device intended to determine a variety of analytes in human body fluids.
Manufacturer(s)
Siemens Healthcare Diagnostics, Inc.