Siemens Healthcare Diagnostics, Inc. recalls Dimension Vista¿ Mass CKMB Isoenzyme Calibrator (MMB CAL) an in vitro diagnostic product for the calibration of Creatine
Recalled October 15, 2014 · FDA recall Z-0024-2015 · Siemens Healthcare Diagnostics
Hazard
Siemens Healthcare Diagnostics has determined that Dimension Vista® Mass CKMB Isoenzyme Calibrator (MMB CAL) shows drift during routine stability monitoring that exceeds our acceptance criteria for drift over the shelf life of this product. Siemens has observed drift from 3% to 12% at MMB concentrations across the assay range. Depending on quality control limits, this drift may not have been d · FDA Class II
Units
6562
Sold at
Worldwided distribution: US (nationwide) and countries of: Australia, Austria, Belgium, Canada, Denmark, Germany, Finland, France, Italy, Japan, Netherlands, New Zealand, Portugal, Saudi Arabia, Slovenia, South Korea, Spain, and Switzerland.
Description
Siemens Healthcare Diagnostics has determined that Dimension Vista® Mass CKMB Isoenzyme Calibrator (MMB CAL) shows drift during routine stability monitoring that exceeds our acceptance criteria for drift over the shelf life of this product. Siemens has observed drift from 3% to 12% at MMB concentrations across the assay range. Depending on quality control limits, this drift may not have been d
Product
Dimension Vista¿ Mass CKMB Isoenzyme Calibrator (MMB CAL) an in vitro diagnostic product for the calibration of Creatine Kinase MB Isoenzyme (MMB) method on the Dimension Vista¿ System.
Manufacturer(s)
Siemens Healthcare Diagnostics, Inc.