RecallWatchUSA

Home / Recalls / Medical Devices

Siemens Healthcare Diagnostics, Inc. recalls Atellica IM Homocysteine Calibrator (2 Pack). Siemens Material Number: 10995498

Recalled June 4, 2025 · FDA recall Z-1821-2025 · Siemens Healthcare Diagnostics

Hazard
Certain lots of Atellica IM Homocysteine Calibrator and ADVIA Centaur HCY Calibrator do not meet the assay Instructions for Use (IFU) reference interval due to calibrator instability resulting in increased positive bias. · FDA Class II
Units
2,212 packs (4,424 units) - 448 packs US, 1,764 packs OUS
Sold at
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KY, MA, MD, ME, MI, MO, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TX, UT, VA, VT, WI, WY and the countries of Austria, Belgium, Croatia, Curacao,¿St¿Eus, Czech¿Republic, Denmark, Egypt, Finland, France, Georgia, Germany, Greece, Hungary, Iran, Ireland, Israel, Italy, Jordan, Kuwait, Latvia, Lithuania, Nepal, Netherlands, Norway, Pakistan, Poland, Portugal, Romania, Serbia, Slovakia, Spain, Sweden, Switzerland, Turkey, Turkmenistan, U.A.E., United¿Kingdom, Vatikan City, Zaire.

Description

Certain lots of Atellica IM Homocysteine Calibrator and ADVIA Centaur HCY Calibrator do not meet the assay Instructions for Use (IFU) reference interval due to calibrator instability resulting in increased positive bias.

Product
Atellica IM Homocysteine Calibrator (2 Pack). Siemens Material Number: 10995498
Manufacturer(s)
Siemens Healthcare Diagnostics, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov