Siemens Healthcare Diagnostics, Inc. recalls Atellica CH ¿2-Microglobulin (B2M) -M) Assay for in vitro diagnostic use in the quantitative determination of ¿2-microgl
Recalled January 5, 2022 · FDA recall Z-0440-2022 · Siemens Healthcare Diagnostics
Hazard
¿2 Microglobulin (B2M) Reagent Lot 211446 Bias with Quality Control and Patient Sample Results. Positive bias on patient samples of up to +20.9% at a B2M concentration of approximately 1.5 mg/L and a negative bias on patient samples of up to -10.5% at a B2M concentration of approximately 15 mg/L · FDA Class II
Units
US: 46 units OUS: 794 units
Sold at
Worldwide distribution - US Nationwide distribution in the states of CA, CO, MA, MD, MS, NC, NE, NH, OH, VA, WI and the countries of Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Czech Republic, Denmark, Ecuador, France, Germany, Greece, India, Iran, Israel, Italy, Japan, Mexico, Morocco, Oman, P.R. China, Pakistan, Philippines, Portugal, Slovakia, South Africa, Spain, Switzerland, Turkey, United Kingdom, Uruguay.
Description
¿2 Microglobulin (B2M) Reagent Lot 211446 Bias with Quality Control and Patient Sample Results. Positive bias on patient samples of up to +20.9% at a B2M concentration of approximately 1.5 mg/L and a negative bias on patient samples of up to -10.5% at a B2M concentration of approximately 15 mg/L
Product
Atellica CH ¿2-Microglobulin (B2M) -M) Assay for in vitro diagnostic use in the quantitative determination of ¿2-microglobulin in human serum and plasma Siemens Material Number (SMN): 11097635