RecallWatchUSA

Home / Recalls / Medical Devices

Siemens Healthcare Diagnostics recalls IMMULITE 2000 /IMMULITE 2000 XPi Systems TIE Total IgE, REF L2KIE2, SMN 10380873, and L2KIE6, SMN 10380872, IVD --- Siem

Recalled May 7, 2014 · FDA recall Z-1518-2014 · Siemens Healthcare Diagnostics

Hazard
Siemens confirmed an overall average positive bias of 23% against the WHO 2nd IRP 75/502 with the IMMULITE Systems Total IgE assays. · FDA Class II
Units
18,489 units
Sold at
Worldwide distribution.

Description

Siemens confirmed an overall average positive bias of 23% against the WHO 2nd IRP 75/502 with the IMMULITE Systems Total IgE assays.

Product
IMMULITE 2000 /IMMULITE 2000 XPi Systems TIE Total IgE, REF L2KIE2, SMN 10380873, and L2KIE6, SMN 10380872, IVD --- Siemens Healthcare Diagnostics Products Ltd. For in vitro diagnostic use with the IMMULITE, IMMULITE 1000, and IMMULITE 2000/ IMMULITE 2000 XPi Systems Analyzers - for the quantitative measurement of Total Immunoglobulin type E (IgE) in serum.
Manufacturer(s)
Siemens Healthcare Diagnostics

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Siemens Healthcare Diagnostics recalls IMMULITE 2000 /IMMULITE 2000 XPi Systems TIE Total IgE, REF L2KIE2, SMN 10380873, and L2KIE6, SMN 10380872, IVD --- Siem | RecallWatchUSA