Siemens Healthcare Diagnostics recalls ADVIA Centaur XP Immunoassay system, an automated in vitro diagnostic analyzer. Catalog # 078-A011-03 - ADVIA Centaur XP
Recalled May 14, 2014 · FDA recall Z-1553-2014 · Siemens Healthcare Diagnostics
Hazard
Instrument Cover Gas Spring failures. Over time, the gas spring may lose its effectiveness and fail to support the cover in its full or partially open position. This may lead to the cover falling. · FDA Class II
Units
5,900 units total (1,679 domestically & 4,221 internationally)
Sold at
Nationwide and internationally
Description
Instrument Cover Gas Spring failures. Over time, the gas spring may lose its effectiveness and fail to support the cover in its full or partially open position. This may lead to the cover falling.
Product
ADVIA Centaur XP Immunoassay system, an automated in vitro diagnostic analyzer. Catalog # 078-A011-03 - ADVIA Centaur XP, Catalog # 078-A011R03 - ADVIA Centaur XP refurb.
Manufacturer(s)
Siemens Healthcare Diagnostics
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →