Home / Recalls / Medical Devices
SEASPINE ORTHOPEDICS CORPORATION recalls Orthofix NorthStar OCT Navigation Surgical Technique Guide, and instruments: Navigation Driver, REF: PC2-400012; Naviga
Recalled December 24, 2025 · FDA recall Z-0937-2026 · SEASPINE ORTHOPEDICS
Hazard
The NorthStar OCT Navigation Surgical Technique Guide (STG) incorrectly identified both Medtronic s Solera and Medtronic s Infinity system toolcards as compatible with the NorthStar OCT Navigation Instruments. However, the NorthStar OCT Navigation Instruments are compatible for use with Medtronic s Solera toolcards only. · FDA Class II
Units
310
Sold at
US: CT, TX, NV, CO, CA, WA, NJ, OR, IN, MI, FL, PA, IL, VA, AL, RI, MO, AZ, MS, SC, UT, NY, OK, SD. OUS: Australia, Mexico, Chile
Description
The NorthStar OCT Navigation Surgical Technique Guide (STG) incorrectly identified both Medtronic s Solera and Medtronic s Infinity system toolcards as compatible with the NorthStar OCT Navigation Instruments. However, the NorthStar OCT Navigation Instruments are compatible for use with Medtronic s Solera toolcards only.
Product
Orthofix NorthStar OCT Navigation Surgical Technique Guide, and instruments: Navigation Driver, REF: PC2-400012; Navigation Straight Bone Probe, REF: PC2-400013; Navigation, 3.5mm Drill, REF: PC2-400135; Navigation, 4.0mm Drill, REF: PC2-400140; Navigation, 4.5mm Drill, PC2-400145; Navigation, 5.5mm Drill, REF: PC2-400155; Navigation, 3.5mm Tap, REF: PC2-400235; Navigation, 4.0mm Tap, REF: PC2-400240; Navigation, 4.5mm Tap, REF: PC2-400245; Navigation, 4.5mm Dual Lead Tap, REF: PC2-400345; Navigation, 5.5mm Dual Lead Tap, REF: PC2-400355
Manufacturer(s)
SEASPINE ORTHOPEDICS CORPORATION
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →