RecallWatchUSA

Home / Recalls / Medical Devices

Richard Wolf GmbH recalls FLEX. GRASP. FORCEPS 5FR WL 550MM, Product Number 8735.685

Recalled May 3, 2023 · FDA recall Z-1424-2023 · Richard Wolf

Hazard
Potentially unable to open forceps jaws when the endoscope shaft is bent by about 90 degrees. If, in addition, the proximal forceps shaft is bent, the forceps can no longer be opened or restrictions begin even with a slight bending of the endoscope. · FDA Class II
Units
319 units
Sold at
Worldwide distribution - US Nationwide and the countries of United Arab Emirates (AE), Argentina (AR), Austria (AT), Belgium (BE), BO, Brazil (BR), Canada (CA), Chile (CL), Czech Republic (CZ), Germany (DE), Denmark (DK), Ecuador (EC), Spain (ES), Finland (FI), France (FR), Greece (GR), Hong Kong (HK), Croatia (HR), Ireland (IE), India (IN), Italy (IT), Jordan (JO), Japan (JP), South Korea (KR), Kuwait (KW), Kazakhstan (KZ), LB, Lithuania (LT), Marzocco (MA), Mexico (MX), Netherlands (NL), Norway (NO), New Zealand (NZ), Oman (OM), Pakistan (PK), Poland (PL), Palestine (PS), Portugal (PT), Qatar (QA), Romania (RO), Serbia (RS), Russian Federation (RU), Saudi Arabia (SA), Slovenia (SI), Slovakia (SK), Thailand (TH), Turkey (TR).

Description

Potentially unable to open forceps jaws when the endoscope shaft is bent by about 90 degrees. If, in addition, the proximal forceps shaft is bent, the forceps can no longer be opened or restrictions begin even with a slight bending of the endoscope.

Product
FLEX. GRASP. FORCEPS 5FR WL 550MM, Product Number 8735.685
Manufacturer(s)
Richard Wolf GmbH
Made in
Germany

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov